Selecting the Direct, Step or Preset generation mode, the final working screen shows the dry well temperature at the bottom (together with the temperature set- point, the generation mode and the status arrow symbol) and the contact state at the top (open/close). An audio signal indicates the change of state.
SWITCH |
|
150.00 | DIRECT |
BLK |
|
148.00 | °C |
To enable the Switch Test with the indication of open, close and calculated hysteresis values, the Ramp generation mode must first be selected.
SWITCH |
|
|
OPENED AT: | 145.50 | °C |
CLOSED AT: | 148.00 | °C |
HIST: | 2.50 | °C |
150.00 |
|
|
BLK: | 148.00 °C |
|
The switch test function can be rapidly recalled from any working screen by pressing the # and +/- keys. Press these keys again to disable Switch Test and to recall only the dry well temperature measurement.
•Thermocouple Testing
The
From the Menu Selection Screen, select the required dry well temperature reference into the REF. T display, move the selection cursor on the MEASURE
section using the keys, select T/C using the keys and press ent.
REF T.
INTERNAL
SEL RTD
MEASURE |
NONE |
T/C |
RTD |
mA |