Nominal Value is the recommended value for the test): pressing the ent key causes the microprocessor to combine the current indications on the display (True Value and Effective Reading).
•Calibration of the resistance (Ohm) input ranges (DBC-TC only)
As for the millivolt input range, also the resistance measurement channel has different ranges capability, depending on the RTD measurement available.
CALIBRATION |
Ohms INPUT |
1 KOhms INPUT |
From the CALIBRATION menu select the required Ohms input range to be
calibrated, using the keys, and press ent.
Ohms INPUT
Nominal value:
300Ω
True value: 300.00Ω
Reading: 298.00Ω
Press ENT or CE ...
Using a certified and high accuracy Ohms simulator, apply the required value to the related electrical connection on the front panel of the instrument (the nominal value is the recommended value for the test).
Pressing the ent key causes the microprocessor to combine the resistance measure on the display (true value and effective reading).
52 | K400 Issue No. 1 |