GE K400 user manual Calibration of the resistance Ohm input ranges DBC-TC only

Page 62

Nominal Value is the recommended value for the test): pressing the ent key causes the microprocessor to combine the current indications on the display (True Value and Effective Reading).

Calibration of the resistance (Ohm) input ranges (DBC-TC only)

As for the millivolt input range, also the resistance measurement channel has different ranges capability, depending on the RTD measurement available.

CALIBRATION

Ohms INPUT

1 KOhms INPUT

From the CALIBRATION menu select the required Ohms input range to be

calibrated, using the keys, and press ent.

Ohms INPUT

Nominal value:

300

True value: 300.00

Reading: 298.00

Press ENT or CE ...

Using a certified and high accuracy Ohms simulator, apply the required value to the related electrical connection on the front panel of the instrument (the nominal value is the recommended value for the test).

Pressing the ent key causes the microprocessor to combine the resistance measure on the display (true value and effective reading).

52

K400 Issue No. 1

Image 62
Contents Druck DBC 150/650 Series General Electric Company. All rights reserved Maintenance SafetyPower isolation Technical AdviceSymbols Traceability and Calibration ReportSafety Instructions AbbreviationsTable of Contents CAL Electrical Input Specification DBC-TC models Standard Accessories DBC-TC series Optional Accessories DBC-TC seriesStandard Accessories DBC-TS series Optional Accessories DBC-TS seriesIntroduction TS Temperature SourceTC Temperature Calibrator Automatic Calibration Parts IdentificationDBC-TS Temperature Probes InstallationPower Supply Important Note Use of screw connections for T/C wires Electrical Connections Connection of a Thermocouple sensorConnection of a 3-wire RTD Connection of a 4-wire RTDOperating procedure GeneralCalibration with external reference probe External probe connections OPERATI0N Numeric keys +/- keyKey Arrow keysWorking screen Menu selection screenSystem Set-up Screen Help/Info screen Self-test screenFunction Modes Reading of the generated temperature Dual display generated temperature and an input test optionExternal reference probe Temperature Reference SelectionInternal reference probe SEL RTD System Set-up Language setting Temperature unit settingDigital interface settings Key-padaudio signalPrinter settings CalibrationProtection code Testing Modes Direct mode 150.00 CStep mode Fixed step selection19.00 C Ramp mode 50.00 CPreset values mode Switch Test Measuring an InputThermocouple Testing 148.00 CSetting Cold Junction Compensation InternalCAL Mode DirectAbsolute SpanReading RTD Measurements Pt-100 Using the keys, select the required method and press ent Reading Current Measurements CAL Voltage Measurements high level CAL Volt Voltage Measurements low level Resistance Measurement DBC Series Serial Commands Protocol Serial CommunicationTS/ TC Read AaOM 1? Preset mode French German Italian Portuguese Spanish Write Par = floating point Calibration of the standard Reference Probe Input CalibrationImportant Note REF RTD Calibration of the millivolt input ranges DBC-TC only Calibration of the Volt input range DBC-TC only Calibration of the mA input range DBC-TC onlyCalibration of the resistance Ohm input ranges DBC-TC only Fault Finding Maintenance Cleaning InstructionsCommon specification Technical SpecificationStandard Specification Electrical Input Specifications DBC-TC models Input Range Resolution Accuracy Days Year