GE K400 user manual Preset values mode

Page 36

Preset values mode

To set-up temperature generation from the standard working screen, using PRESET values, the user must assign a temperature set-point value to a specific numeric key on the key-pad. From the Direct/Step/Ramp working screen, press ce

since the setting menu is already open. Select PRESET using the keys and press ent.

SET MODE

DIRECT

PRESET

STEPS

RAMP

PRESET

# 1

=

0.00

°C

# 2

=

0.00

°C

# 3

=

0.00

°C

# 4

=

0.00

°C

# 5

=

0.00

°C

The next display allows a specific temperature value to be assigned to each of the

numeric keys (1 to 5) of the key-pad. Use the keys and the numeric key-pad to set the values. Press ent to leave the set-up display and to open the working screen.

Pressing the # key followed by one of the numeric keys 1- 5 recalls the set-point assigned to the numeric key.

Note:

Ramp, Step and Preset functions can also be set starting from the Menu Selection Screen. In an active standard working screen, press the menu

key to open the Menu Selection Screen and use the keys to select the required temperature reference from the REF. T display (see next section), and press ent. The following screen allows set-up of the Ramp, Step and Preset parameters.

26

K400 Issue No. 1

Image 36
Contents Druck DBC 150/650 Series General Electric Company. All rights reserved Safety Power isolationMaintenance Technical AdviceSymbols Traceability and Calibration ReportSafety Instructions AbbreviationsTable of Contents CAL Electrical Input Specification DBC-TC models Standard Accessories DBC-TC series Optional Accessories DBC-TC seriesStandard Accessories DBC-TS series Optional Accessories DBC-TS seriesTS Temperature Source TC Temperature CalibratorIntroduction Automatic Calibration Parts IdentificationDBC-TS Installation Power SupplyTemperature Probes Important Note Use of screw connections for T/C wires Electrical Connections Connection of a Thermocouple sensorConnection of a 3-wire RTD Connection of a 4-wire RTDOperating procedure GeneralCalibration with external reference probe External probe connections OPERATI0N +/- key KeyNumeric keys Arrow keysMenu selection screen System Set-up ScreenWorking screen Help/Info screen Self-test screenFunction Modes Reading of the generated temperature Dual display generated temperature and an input test optionTemperature Reference Selection Internal reference probeExternal reference probe SEL RTD System Set-up Language setting Temperature unit settingDigital interface settings Key-padaudio signalPrinter settings CalibrationProtection code Testing Modes Direct mode 150.00 C Step mode Fixed step selection19.00 C Ramp mode 50.00 CPreset values mode Switch Test Measuring an InputThermocouple Testing 148.00 CSetting Cold Junction Compensation InternalCAL Mode DirectAbsolute SpanReading RTD Measurements Pt-100 Using the keys, select the required method and press ent Reading Current Measurements CAL Voltage Measurements high level CAL Volt Voltage Measurements low level Resistance Measurement DBC Series Serial Commands Protocol Serial CommunicationTS/ TC Read AaOM 1? Preset mode French German Italian Portuguese Spanish Write Par = floating point Calibration Important NoteCalibration of the standard Reference Probe Input REF RTD Calibration of the millivolt input ranges DBC-TC only Calibration of the Volt input range DBC-TC only Calibration of the mA input range DBC-TC onlyCalibration of the resistance Ohm input ranges DBC-TC only Fault Finding Maintenance Cleaning InstructionsTechnical Specification Standard SpecificationCommon specification Electrical Input Specifications DBC-TC models Input Range Resolution Accuracy Days Year