GE K400 user manual 50.00 C, Ramp mode

Page 35

Ramp mode

The DBC series can generate temperature using programmable temperature ramp function. To set-up a ramp function from a Direct or Step working screen, press the

#and 0 keys. The next display allows set-up of the ramp temperature limits and the required temperature gradient (°C/minute), to be used during the generation (maximum capability 10.0°C/minute).

RAMP

T min:

0.00

°C

T max:

100.00

°C

dt/ dT:

1.00

°C/ min

Move the cursor in vertical direction using the keys and, using the numeric key- pad, set the start/finish ramp temperature values (T min and T max) and the required rate of change for the regulator. Pressing the ent key causes the working screen to be activated. Use the key to start ramp generation, starting at the actual dry well temperature and finishing at the T max value. Use the key to start a negative ramp function (actual dry well temperature to T min). The display shows the ramp direction, together with a corresponding heating/cooling symbol.

60.00

BLK:

50.00 °C

To stop the ramp, press the key. Press the keys to restart. The status bar

shows the ramp symbol, the temperature set-point, the symbols and a HOLD message (if the ramp has been stopped).

K400 Issue No. 1

25

Image 35
Contents Druck DBC 150/650 Series General Electric Company. All rights reserved Technical Advice SafetyPower isolation MaintenanceTraceability and Calibration Report SymbolsAbbreviations Safety InstructionsTable of Contents CAL Electrical Input Specification DBC-TC models Optional Accessories DBC-TC series Standard Accessories DBC-TC seriesOptional Accessories DBC-TS series Standard Accessories DBC-TS seriesIntroduction TS Temperature SourceTC Temperature Calibrator Parts Identification Automatic CalibrationDBC-TS Temperature Probes InstallationPower Supply Important Note Electrical Connections Connection of a Thermocouple sensor Use of screw connections for T/C wiresConnection of a 4-wire RTD Connection of a 3-wire RTDGeneral Operating procedureCalibration with external reference probe External probe connections OPERATI0N Arrow keys +/- keyKey Numeric keysWorking screen Menu selection screenSystem Set-up Screen Self-test screen Help/Info screenDual display generated temperature and an input test option Function Modes Reading of the generated temperatureExternal reference probe Temperature Reference SelectionInternal reference probe SEL RTD Temperature unit setting System Set-up Language settingKey-padaudio signal Digital interface settingsCalibration Printer settingsProtection code 150.00 C Testing Modes Direct modeFixed step selection Step mode19.00 C 50.00 C Ramp modePreset values mode Measuring an Input Switch Test148.00 C Thermocouple TestingInternal Setting Cold Junction CompensationDirect CAL ModeSpan AbsoluteReading RTD Measurements Pt-100 Using the keys, select the required method and press ent Reading Current Measurements CAL Voltage Measurements high level CAL Volt Voltage Measurements low level Resistance Measurement Serial Communication DBC Series Serial Commands ProtocolTS/ TC Read AaOM 1? Preset mode French German Italian Portuguese Spanish Write Par = floating point Calibration of the standard Reference Probe Input CalibrationImportant Note REF RTD Calibration of the millivolt input ranges DBC-TC only Calibration of the mA input range DBC-TC only Calibration of the Volt input range DBC-TC onlyCalibration of the resistance Ohm input ranges DBC-TC only Fault Finding Cleaning Instructions MaintenanceCommon specification Technical SpecificationStandard Specification Input Range Resolution Accuracy Days Year Electrical Input Specifications DBC-TC models