Using a decade resistance box, generate the required value (to the Ref. Probe connection on the electrical panel) and press ent. The microprocessor combines the two values automatically. The nominal value is the value recommended for the test.
REF. RTD INPUT
Nominal value:
300Ω
True value: | 300.00 |
Reading: | 298.50 |
Press ENT or CE ...
•Calibration of the external Reference Probe
By selecting the REF RTD from the calibration menu and pressing ent, a new display opens to set all the parameters needed for the connected Reference RTD (Tag number, R0 values, coefficients).
REF RTD
Tag:
R0:100.00Ω
A:
B:
C:
Move the cursor with the keys and set the value with the numeric
•Calibration of the internal dry well Sensor
Note: Starting this procedure after at least 1 hour of
CALIBRATION | |
SET 1: | 50.00 |
REF 1: | 50.00 |
SET 2: | 300.00 |
REF 2: | 300.00 |
SET 3: | 400.00 |
K400 Issue No. 1 | 49 |