Agilent Technologies 11970-90031 manual Save the average conversion-loss value

Page 32

Operation

Using the Mixers with the 856X Series Spectrum Analyzers

In this example, we’ll look at U-band, which ranges from 40 GHz to 60 GHz, as shown in Figure 2-9. The LOCK HARMONIC function “locks” the spectrum analyzer in that band, ensuring that the spectrum analyzer sweeps only the chosen band. LOCK ON OFF is automatically set to ON when FULL BAND is pressed.

Figure 2-9 Select the band of interest.

Operation

Save the average conversion-loss value

4.Default conversion-loss values that are stored in the analyzer for each frequency band are listed in Table 2-1. These values approximate the values for the Agilent 11970 series mixers. Other conversion-loss values may be entered into the spectrum analyzer in two ways. The first method lets you to save the average conversion-loss value for the entire band using AVERAGE CNV LOSS. To activate this function:

a.Press AUX CTRL, EXTERNAL MIXER, AMPTD CORRECT, AVERAGE CNV LOSS.

b.Enter the appropriate average conversion-loss value. On Agilent 11970 Series harmonic mixers, these values are printed on the mixer. The U-band mixer used for this example had an average conversion loss of 23.5 dB, as shown in Figure 2-10.

32

Chapter 2

Image 32
Contents Manufacturing Part Number OctoberCertification Limitation of Warranty Page Contents Service General Information Introduction Serial NumbersSeries Harmonic Mixers Mixer Connection Kit, Option Specifications Waveguide Connector Screws11970 Series Specifications LO Harmonic Number LO Input Frequency Range Model RF Frequency Range 11970 Supplemental Characteristics Specifications Operation Operating Precautions IntroductionGetting Started Mixer ConnectionsGetting Started Loading Data Viewing DataSet up the equipment Harmonic Mixer SetupAmplitude Calibration Manually Entering Conversion-Loss DataSignal Identification Chapter Spectrum Analyzer Amplitude Calibration Signal Identification AYZ Using the Mixers with the 856X Series Spectrum Analyzers External Mixer SetupSelect the Frequency Band External Mixer Frequency RangesSave the average conversion-loss value Select the band of interestSignal Identification 10 Store and correct for conversion lossIdentify signals with the frequency-shift method 11 Signal Responses Produced by a 50 GHz Signal in U Band12 Response for Invalid Signals Identify signals in wide frequency spans 14 SIG ID AT MKR Performed on an Image Signal15 SIG ID AT MKR Performed on a True Signal Equipment Connection Using the Mixers with MMS AnalyzersPreliminary Operation Amplitude Calibration for a Single Frequency OperationBand Selection External Mixing Bands and Associated Mixing Harmonics Band Harmonic Number N Frequency Range GHzSets which active input port will be selected when the Instr Frequency Ranges and Corresponding Harmonic Numbers Conversion Loss Versus Frequency Correction For further information, consult the Agilent 70900B Using the Mixers with the 8566B Spectrum Analyzer 17External Mixer ConnectionProcedure Amplitude CalibrationSignal Identification 19 Sample Mixer Calibration Table Performance Tests Test Record Recommended Test Equipment for 11907K Performance Test ProceduresDescription Recommended Test Equipment for 11970A Recommended Test Equipment for 11970Q Recommended Test Equipment for 11970U Recommended Test Equipment for Recommended Test Equipment for 11970W Specifications Conversion Loss and Frequency ResponseDescription 11970V 50.0 GHz 11970W 75.0 GHz For 11970V or W Then press Center Frequency Conversion Loss and Frequency Response Conversion Loss and Frequency Response Performance Test Setups 1 Performance Test Setups 2 Conversion Loss and Frequency Response Test Record Conversion Loss and Frequency Response Test Record Specification Average Noise Level TestAverage Noise Level Test For 11970V or W Average Noise Level Test Average Noise Test Record Average Noise LevelService Maintenance RepairsReplaceable Parts Circuit Description11970 Series Mixer Schematic Diagram Accessories and Replaceable Parts Part Number DescriptionAgilent Technologies Sales and Service Offices European Field OperationsMaintenance