Operation
Using the Mixers with MMS Analyzers
Operation
Band Selection
Use the following key sequence to enter the external mixing mode and to select the desired band of operation:
MENU
Select Input - choose external mixer input, for example: “IN 2 EM 70910A”
State
ext mixer
fulband KAQUVE
Press the softkey for the desired band.
NOTE | The above band selection key sequence must be used to enter the external |
| mixing band of operation. Do not enter in the center frequency of the band |
| directly. |
|
|
Amplitude Calibration for a Single Frequency
Enter the conversion loss at the desired frequency from either the calibration label on the end of the Agilent 11970 or the supplied calibration sheet by using the following key sequence:
State
ext mixer
CONV LOSS
Enter the conversion loss value of the desired frequency, then terminate the entry with dB.
NOTE | The above procedure is a |
| alternative procedure, available by remote programming only, refer to |
| “Conversion Loss Versus Frequency Correction” on page 43. |
|
|