Agilent Technologies 11970-90031 manual Performance Test Procedures, Description

Page 51

Performance Tests

Performance Test Procedures

Performance Test Procedures

Description

Each performance test procedure is contained in a single paragraph. The first entry in each paragraph is the specification for the parameter being measured as described in Table 1-1This is followed by a general description of the test and any special instructions or problem areas. Appropriate test setup illustrations are included in this section and are referenced in the procedures. You MUST do the tests, and the steps within each test, in the order they are given.

Table 3-1. Recommended Test Equipment for 11907K

Instrument

Critical Specifications

Recommended Model

 

 

 

Spectrum Analyzer

LO and IF ranges compatible with mixer

8566B

Synthesized Sweeper

Frequency: 18 to 26.5 GHz

8340A

 

Output Level: >-10 dBm

 

Amplifier

Output Level: >18 dBm leveled

11975A

 

Frequency Range: 3 to 4.5 GHz

 

Power Meter

Compatible with Power Sensor

436A

Power Sensor

SWR: <1.3

8485A

Directional Coupler1

Coupling: 10 dB

K752C

 

Directivity: >40 dB

 

 

Primary Arm SWR: <1.05

 

 

Auxiliary Arm SWR: <1.2

 

Isolator

Insertion Loss: <1.5 dB

P/N 0960-0081

 

Isolation: >20 dB

 

 

SWR: <1.2

 

Adapter (2 required)

3.5 mm female to WR-42

K281C

 

SWR: <1.1

 

Cables (3 required)

Connectors: SMA male

P/N 5061-5458

Cable

Connectors: SMA

P/N 8120-4396

 

Loss: <1.0 dB @ 20 GHz

 

 

 

 

1. Calibration data for the coupling ratio between the output arm and the auxiliary

arm is necessary for accurate measurements.Performance Tests

Chapter 3

51

Image 51
Contents October Manufacturing Part NumberCertification Limitation of Warranty Page Contents Service General Information Serial Numbers IntroductionSeries Harmonic Mixers Mixer Connection Kit, Option Waveguide Connector Screws Specifications11970 Series Specifications LO Harmonic Number LO Input Frequency Range Model RF Frequency Range 11970 Supplemental Characteristics Specifications Operation Introduction Operating PrecautionsMixer Connections Getting StartedGetting Started Viewing Data Loading DataHarmonic Mixer Setup Set up the equipmentManually Entering Conversion-Loss Data Amplitude CalibrationSignal Identification Chapter Spectrum Analyzer Amplitude Calibration Signal Identification AYZ External Mixer Setup Using the Mixers with the 856X Series Spectrum AnalyzersExternal Mixer Frequency Ranges Select the Frequency BandSelect the band of interest Save the average conversion-loss value10 Store and correct for conversion loss Signal Identification11 Signal Responses Produced by a 50 GHz Signal in U Band Identify signals with the frequency-shift method12 Response for Invalid Signals 14 SIG ID AT MKR Performed on an Image Signal Identify signals in wide frequency spans15 SIG ID AT MKR Performed on a True Signal Using the Mixers with MMS Analyzers Preliminary OperationEquipment Connection Operation Band SelectionAmplitude Calibration for a Single Frequency Band Harmonic Number N Frequency Range GHz External Mixing Bands and Associated Mixing HarmonicsSets which active input port will be selected when the Instr Frequency Ranges and Corresponding Harmonic Numbers Conversion Loss Versus Frequency Correction For further information, consult the Agilent 70900B 17External Mixer Connection Using the Mixers with the 8566B Spectrum AnalyzerAmplitude Calibration ProcedureSignal Identification 19 Sample Mixer Calibration Table Performance Tests Test Record Performance Test Procedures DescriptionRecommended Test Equipment for 11907K Recommended Test Equipment for 11970A Recommended Test Equipment for 11970Q Recommended Test Equipment for 11970U Recommended Test Equipment for Recommended Test Equipment for 11970W Conversion Loss and Frequency Response SpecificationsDescription 11970V 50.0 GHz 11970W 75.0 GHz For 11970V or W Then press Center Frequency Conversion Loss and Frequency Response Conversion Loss and Frequency Response Performance Test Setups 1 Performance Test Setups 2 Conversion Loss and Frequency Response Test Record Conversion Loss and Frequency Response Test Record Average Noise Level Test SpecificationAverage Noise Level Test For 11970V or W Average Noise Level Test Average Noise Level Average Noise Test RecordService Circuit Description MaintenanceRepairs Replaceable Parts11970 Series Mixer Schematic Diagram Part Number Description Accessories and Replaceable PartsEuropean Field Operations Agilent Technologies Sales and Service OfficesMaintenance