Agilent Technologies 11970-90031 manual Amplitude Calibration, Procedure

Page 46

Operation

Operation

Using the Mixers with the 8566B Spectrum Analyzer

“Harmonic 6L”, where “L” stands for locked, or you could press Shift+z to manually harmonic lock.

Figure 2-18TYPICAL MILLIMETER SIGNAL DISPLAY

A menu of millimeter measurement functions is accessed by pressing SHIFT, 1, MHz. Even though the spectrum analyzer is in an external mixer band, the 8566B spectrum analyzer retains full operation capability (except for use of the RF input attenuator and preselector).

Amplitude Calibration

Procedure

Each 11970 series harmonic mixer is characterized at the factory and comes with a calibration chart (See Figure 2-19). The horizontal axis of the chart gives the frequency range of the mixer and the two vertical axis scales are calibrated in conversion loss and in reference level offset.

To correct the amplitude calibration of the spectrum analyzer to account for the conversion loss of the harmonic mixer, press SHIFT, 1, MHz, which accesses the millimeter menu. Then press SHIFT, 5, MHz to access conversion loss entry and key in the appropriate value on the spectrum analyzer as shown on the mixer calibration chart.

For measurements that span the entire band, choose an average value of conversion loss from the conversion loss chart. For those spectrum analyzers with date codes earlier than 13.8.86, the conversion loss is changed using the reference level function. Press SHIFT, then REFERENCE LEVEL, then key in the reference level offset shown on the right side of the mixer calibration chart.

46

Chapter 2

Image 46
Contents Manufacturing Part Number OctoberCertification Limitation of Warranty Page Contents Service General Information Introduction Serial NumbersSeries Harmonic Mixers Mixer Connection Kit, Option Specifications Waveguide Connector Screws11970 Series Specifications LO Harmonic Number LO Input Frequency Range Model RF Frequency Range 11970 Supplemental Characteristics Specifications Operation Operating Precautions IntroductionGetting Started Mixer ConnectionsGetting Started Loading Data Viewing DataSet up the equipment Harmonic Mixer SetupAmplitude Calibration Manually Entering Conversion-Loss DataSignal Identification Chapter Spectrum Analyzer Amplitude Calibration Signal Identification AYZ Using the Mixers with the 856X Series Spectrum Analyzers External Mixer SetupSelect the Frequency Band External Mixer Frequency RangesSave the average conversion-loss value Select the band of interestSignal Identification 10 Store and correct for conversion lossIdentify signals with the frequency-shift method 11 Signal Responses Produced by a 50 GHz Signal in U Band12 Response for Invalid Signals Identify signals in wide frequency spans 14 SIG ID AT MKR Performed on an Image Signal15 SIG ID AT MKR Performed on a True Signal Preliminary Operation Using the Mixers with MMS AnalyzersEquipment Connection Band Selection OperationAmplitude Calibration for a Single Frequency External Mixing Bands and Associated Mixing Harmonics Band Harmonic Number N Frequency Range GHzSets which active input port will be selected when the Instr Frequency Ranges and Corresponding Harmonic Numbers Conversion Loss Versus Frequency Correction For further information, consult the Agilent 70900B Using the Mixers with the 8566B Spectrum Analyzer 17External Mixer ConnectionProcedure Amplitude CalibrationSignal Identification 19 Sample Mixer Calibration Table Performance Tests Test Record Description Performance Test ProceduresRecommended Test Equipment for 11907K Recommended Test Equipment for 11970A Recommended Test Equipment for 11970Q Recommended Test Equipment for 11970U Recommended Test Equipment for Recommended Test Equipment for 11970W Specifications Conversion Loss and Frequency ResponseDescription 11970V 50.0 GHz 11970W 75.0 GHz For 11970V or W Then press Center Frequency Conversion Loss and Frequency Response Conversion Loss and Frequency Response Performance Test Setups 1 Performance Test Setups 2 Conversion Loss and Frequency Response Test Record Conversion Loss and Frequency Response Test Record Specification Average Noise Level TestAverage Noise Level Test For 11970V or W Average Noise Level Test Average Noise Test Record Average Noise LevelService Replaceable Parts MaintenanceRepairs Circuit Description11970 Series Mixer Schematic Diagram Accessories and Replaceable Parts Part Number DescriptionAgilent Technologies Sales and Service Offices European Field OperationsMaintenance