Agilent Technologies 11970-90031 manual Conversion Loss and Frequency Response

Page 63

Performance Tests

Conversion Loss and Frequency Response

NOTE

The conversion loss indicated on the mixer calibration label includes the loss

 

in the IF cable. If other than the specified cable is used, then the loss in that

 

cable must be compensated for when making amplitude measurements.

 

 

18.Increment the frequency of the signal generator 500 MHz higher.

19.Press DATA [], then PEAK SEARCH and MKRCF on the spectrum analyzer.

20.Repeat steps 15 through 18 until the appropriate frequency listed below is reached.

11970K: 26.5 GHz

11970A: 40.0 GHz

11970Q: 50.0 GHz

1197OU: 60.0 GHz

11970V: 75.0 GHz

11970W: 110.0 GHz

21.Repeat steps 1 through 19 for LO inputs to the mixer of 14.5 dBm, 16.0 dBm, and

18.0dBm. In step 5, measure each of these levels at the end of the cable normally connected to the mixer LO input.

22.Frequency response is the difference between the maximum and minimum conversion losses recorded in Table 3-7. For LO power levels between 14.5 and 16.0 dBm this difference must be less than:

3.8dB (for 11970K, 11970A, 11970Q or 1197OU)

4.2dB (for 11970V)

6.0dB (for 11970W)

For LO power levels between 14.0 and 18.0 dBm, the difference must be less than:

5.6dB (for 11970K, 11970A, 11970Q, 11970U or 11970V)

8.0dB (for 11970W)

23.Maximum conversion loss must not exceed the following limits:

For an LO input power between 14.0 and 18.0 dBm.

11970K: 24 dB

11970A: 26 dB

11970Q: 28 dB

11970U: 28 dB

11970V: 40 dB

11970W: 46 dB

Performance Tests

Chapter 3

63

Image 63
Contents October Manufacturing Part NumberCertification Limitation of Warranty Page Contents Service General Information Serial Numbers IntroductionSeries Harmonic Mixers Mixer Connection Kit, Option Waveguide Connector Screws Specifications11970 Series Specifications LO Harmonic Number LO Input Frequency Range Model RF Frequency Range 11970 Supplemental Characteristics Specifications Operation Introduction Operating PrecautionsMixer Connections Getting StartedGetting Started Viewing Data Loading DataHarmonic Mixer Setup Set up the equipmentManually Entering Conversion-Loss Data Amplitude CalibrationSignal Identification Chapter Spectrum Analyzer Amplitude Calibration Signal Identification AYZ External Mixer Setup Using the Mixers with the 856X Series Spectrum AnalyzersExternal Mixer Frequency Ranges Select the Frequency BandSelect the band of interest Save the average conversion-loss value10 Store and correct for conversion loss Signal Identification11 Signal Responses Produced by a 50 GHz Signal in U Band Identify signals with the frequency-shift method12 Response for Invalid Signals 14 SIG ID AT MKR Performed on an Image Signal Identify signals in wide frequency spans15 SIG ID AT MKR Performed on a True Signal Using the Mixers with MMS Analyzers Preliminary OperationEquipment Connection Operation Band SelectionAmplitude Calibration for a Single Frequency Band Harmonic Number N Frequency Range GHz External Mixing Bands and Associated Mixing HarmonicsSets which active input port will be selected when the Instr Frequency Ranges and Corresponding Harmonic Numbers Conversion Loss Versus Frequency Correction For further information, consult the Agilent 70900B 17External Mixer Connection Using the Mixers with the 8566B Spectrum AnalyzerAmplitude Calibration ProcedureSignal Identification 19 Sample Mixer Calibration Table Performance Tests Test Record Performance Test Procedures DescriptionRecommended Test Equipment for 11907K Recommended Test Equipment for 11970A Recommended Test Equipment for 11970Q Recommended Test Equipment for 11970U Recommended Test Equipment for Recommended Test Equipment for 11970W Conversion Loss and Frequency Response SpecificationsDescription 11970V 50.0 GHz 11970W 75.0 GHz For 11970V or W Then press Center Frequency Conversion Loss and Frequency Response Conversion Loss and Frequency Response Performance Test Setups 1 Performance Test Setups 2 Conversion Loss and Frequency Response Test Record Conversion Loss and Frequency Response Test Record Average Noise Level Test SpecificationAverage Noise Level Test For 11970V or W Average Noise Level Test Average Noise Level Average Noise Test RecordService Circuit Description MaintenanceRepairs Replaceable Parts11970 Series Mixer Schematic Diagram Part Number Description Accessories and Replaceable PartsEuropean Field Operations Agilent Technologies Sales and Service OfficesMaintenance