Performance Tests

Performance Tests

Conversion Loss and Frequency Response

Description

The frequency response and conversion loss are checked at four LO power levels. A known input power is applied to the input of the mixer. The IF output power is measured on the 8566B Spectrum Analyzer. From these measurements, the conversion loss and frequency response are calculated.

1.Connect an SMA cable from the 1st LO OUTPUT of the spectrum analyzer to the INPUT of the amplifier. Connect a second SMA cable to the OUTPUT of the amplifier.

2.For 11970K: Zero and calibrate the power meter. For 11970A, 11970Q, 11970U, 11970V

or 11970W: Connect the 478A Power Sensor to the power meter sensor cables, then zero the power meter.

3.Set the 8566A/B Spectrum Analyzer controls as follows:

SHIFT, , KSU FREQUENCY SPAN, 0, Hz SHIFT, SWEEP, CONT, KSt CENTER, 2, 4, [.] 5, GHz

CAUTION When you are using a 11975A Amplifier with a 11970 Series Mixer, you MUST set the amplifier rear-panel ALC switch to ON before you connect the amplifier into the test setup. If the ALC switch is left in the OFF position, the amplifier output power is high enough to destroy the mixer diodes.

4.On the 11975A Amplifier, set the rear panel ALC switch to ON, then connect the power sensor to the free end of the cable installed on the OUTPUT connector of the amplifier. Set the power meter Cal Factor to the appropriate value for a frequency of 4 GHz.

5.Adjust the amplifier OUTPUT POWER LEVEL for a reading of 14.0 ± 0.1 dBm on the power meter. Record the LO power in Table 3-7.

6.Set the power meter Cal Factor to 100 percent.

7.Connect the equipment, as shown in Figure 3-1.

CAUTION Make sure the 8349A Amplifier, used in the signal generator system for 11970A, Q and U tests, is set for external leveling before you turn it on. Failure to do so may allow the amplifier output to rise about 20 dBm (high enough to damage the frequency tripler).

8.Set the signal generator for a CW output signal at the frequencies listed below: 11970K: 18.0 GHz

11970A: 26.5 GHz

11970Q: 33.0 GHz

11970U: 40.0 GHz

58

Chapter 3

Image 58
Contents Manufacturing Part Number OctoberCertification Limitation of Warranty Page Contents Service General Information Introduction Serial NumbersSeries Harmonic Mixers Mixer Connection Kit, Option Specifications Waveguide Connector Screws11970 Series Specifications LO Harmonic Number LO Input Frequency Range Model RF Frequency Range 11970 Supplemental Characteristics Specifications Operation Operating Precautions IntroductionGetting Started Mixer ConnectionsGetting Started Loading Data Viewing DataSet up the equipment Harmonic Mixer SetupAmplitude Calibration Manually Entering Conversion-Loss DataSignal Identification Chapter Spectrum Analyzer Amplitude Calibration Signal Identification AYZ Using the Mixers with the 856X Series Spectrum Analyzers External Mixer SetupSelect the Frequency Band External Mixer Frequency RangesSave the average conversion-loss value Select the band of interestSignal Identification 10 Store and correct for conversion lossIdentify signals with the frequency-shift method 11 Signal Responses Produced by a 50 GHz Signal in U Band12 Response for Invalid Signals Identify signals in wide frequency spans 14 SIG ID AT MKR Performed on an Image Signal15 SIG ID AT MKR Performed on a True Signal Preliminary Operation Using the Mixers with MMS AnalyzersEquipment Connection Band Selection OperationAmplitude Calibration for a Single Frequency External Mixing Bands and Associated Mixing Harmonics Band Harmonic Number N Frequency Range GHzSets which active input port will be selected when the Instr Frequency Ranges and Corresponding Harmonic Numbers Conversion Loss Versus Frequency Correction For further information, consult the Agilent 70900B Using the Mixers with the 8566B Spectrum Analyzer 17External Mixer ConnectionProcedure Amplitude CalibrationSignal Identification 19 Sample Mixer Calibration Table Performance Tests Test Record Description Performance Test ProceduresRecommended Test Equipment for 11907K Recommended Test Equipment for 11970A Recommended Test Equipment for 11970Q Recommended Test Equipment for 11970U Recommended Test Equipment for Recommended Test Equipment for 11970W Specifications Conversion Loss and Frequency ResponseDescription 11970V 50.0 GHz 11970W 75.0 GHz For 11970V or W Then press Center Frequency Conversion Loss and Frequency Response Conversion Loss and Frequency Response Performance Test Setups 1 Performance Test Setups 2 Conversion Loss and Frequency Response Test Record Conversion Loss and Frequency Response Test Record Specification Average Noise Level TestAverage Noise Level Test For 11970V or W Average Noise Level Test Average Noise Test Record Average Noise LevelService Replaceable Parts MaintenanceRepairs Circuit Description11970 Series Mixer Schematic Diagram Accessories and Replaceable Parts Part Number DescriptionAgilent Technologies Sales and Service Offices European Field OperationsMaintenance