35040
Operators Manual
Measurement Accuracy: Charge measurement accuracy is warranted to be better than ± (0.20%rdg + 2 counts) over the operating temperature range of 18°C to 28°C for 1 year.
Field Calibration: Field calibrations may be performed via a restricted distribution
Timed Charge Measurements
Range and Resolution: Same as charge measurements.
Measurement Method: The timed charge measurement is derived by monitoring the charge integrated over a time interval that begins with the activation of the timer. The timer automatically stops when the elapsed time is equal to the timer interval setting or the exposure is over. Interpolation is used to determine the charge at an integral number of seconds.
Measurement Accuracy: Same as charge measurements assuming that the rate of charge delivery does not dramatically change within the interpolation window interval.
Current Measurements
Range, Resolution, and Noise:
Full Scale Range | Display Resolution | Noise | |||
(± counts) | |||||
|
|
|
| ||
200.0 | pA | 0.1 | pA | 2 | |
|
|
|
| ||
2.000 | nA | 0.001 nA | 2 | ||
|
|
|
|
| |
20.00 | nA | 0.01 | nA | 1 | |
|
|
|
|
| |
200.0 | nA | 0.1 | nA | 1 | |
|
|
|
| ||
1.000 | μA | 0.001 μA | 1 | ||
|
|
|
|
|
Measurement Method: The basic current measurement is derived by monitoring the change in the integrated charge level over successive
Measurement Accuracy: Current measurement accuracy is ± (0.20%rdg + 2 counts) over the operating temperature range of 18°C to 28°C for 1 year. Since the current measurements are based directly upon charge measurements, an independent current calibration is not required.
Measurement Noise: Current measurement noise is typically 1 to 4 counts peak to peak (depending upon the range) when the input is supplied by a quiet DC current source. When using a 0.6 cc Farmer Probe to measure the output of a typical 60Co source, noise in the current measurement may be on the order of 0.4 R/min peak to peak at 100 R/min. Normal variations in the output exposure rate produced by a linear accelerator may result in excessive noise on the current measurement.
Response Speed: The display of current measurement values will begin within 1 second of the start of the exposure and reach 99% of the final level within 2 seconds.
Average Current Measurements
Range and Resolution:
Full Scale Range | Display Resolution | |
200.00 pA | 0.01 | pA |
|
| |
2.0000 nA | 0.0001 nA | |
|
|
|
20.000 nA | 0.001 | nA |
|
|
|
200.00 nA | 0.01 | nA |
|
|
|