35040
Operators Manual
Table
Quantity | Description | Units | Comments | |
|
|
|
| |
| Air Density |
| Calculated by instrument and depends on front panel | |
ADCF | Correction | None | ||
Temperature and Pressure settings. | ||||
| Factor |
| ||
|
|
| ||
| Charge during |
| Electrical charge delivered to readout in response to | |
QExposure | C | radiation incident upon detector since the last | ||
exposure | ||||
|
| exposure. | ||
|
|
| ||
QAccum | Accumulated | C | The sum of all charge accumulated due to exposures | |
Charge | since the RESET/MEASURE button was last pressed. | |||
|
| |||
| Previous |
| The sum of all accumulated charge not including the | |
QAccum_previous | Accumulated | C | current exposure since the RESET/MEASURE button | |
| Charge |
| was last pressed. | |
QElectrometer | Electrometer | C | Factory calibrated electrical charge. | |
Charge | ||||
|
|
| ||
Q | Charge | C | Calibrated electrical charge. | |
|
|
|
| |
QScaleFactor | Charge Scale | C / C | Calibrates the charge measured by the electrometer. | |
Factor | Viewed on a Test Function screen. | |||
|
| |||
QOffset | Charge | C | The electrical charge measured when no incident | |
Leakage | radiation is present. | |||
|
| |||
I | Current | A | Calibrated electrical current. | |
|
|
|
| |
IAvg | Average | A | The constant electrical current required to accumulate | |
Current | charge Q over time TStart to T. | |||
|
| |||
|
| Ion |
| |
ICCF | Ion Chamber | Chamber | Values are field programmable via Customization | |
Calibration | Calibration | |||
Software. | ||||
| Factor | Units | ||
|
| |||
|
| / A or C |
| |
ILeakage | Leakage | A | The electrical current measured when no incident | |
Current | radiation is present. | |||
|
| |||
IScaleFactor | Current Scale | A / A | Calibrates the current. Viewed on a Test Function | |
Factor | screen. | |||
|
| |||
RateMultiplier | Rate Units | s | Specifies the number of seconds the rate divisor is | |
Multiplier | equal to 1(s), 60(min), 3600(hr). | |||
|
| |||
T | Q Sampling | s | The time corresponding to Q and I. | |
Time | ||||
|
|
| ||
Time | Effective Time | s | The time required to accumulate charge QExposure | |
of Exposure | with a constant electrical current IAvg. | |||
|
| |||
TOffset | QOffset | s | The time corresponding to QOffset and ILeakage. | |
Sampling Time | ||||
|
|
| ||
QTimed | Timed Charge | C | The electrical charge delivered during the timer | |
interval. | ||||
|
|
| ||
| Predicted |
| The total amount of timed charge that should be | |
QTimed_Predicted | C | accumulated at the completion of the timer interval, | ||
Timed Charge | ||||
|
| based on the observed delivery rate. | ||
|
|
|