35040
Operators Manual
ADC
OVER nC OVER pA
OVER min OVER s
ADC
Current Overrange Reduce exposure and Press RESET/MEASURE
Figure 3-28. The Measurement Screen With the Charge and Current Limits Exceeded After the First 60 Seconds
Figure 3-29. The Measurement Overrange Screen
Table
Table 3-1. Overrange Conditions
Condition | Values Overranged |
|
|
charge > 2.4 μC | charge, accumulated charge, dose, accumulated dose, average |
| current, average dose rate, effective exposure time |
accumulated charge > 1.0000 C | accumulated charge, accumulated dose |
|
|
current > 100.0 nA | all except leakage |
|
|
time > 59 min 59 s | effective exposure time |
|
|
3.9.14 Reset Error
Pressing the RESET /MEASURE button while on the Measurement Screen causes the Model 35040 Therapy Dosimeter to reset the electrometer, zero all readings, and restart the measurement process. If an exposure starts during the reset operation, the instrument displays the reset error message (see Figure
ADC
Internal Reset Error
Press RESET/MEASURE
and Repeat Exposure
Figure 3-30. The Measurement Screen Showing a Reset Error
To remove the error condition, end the exposure, and press the RESET/MEASURE button. Wait a few seconds and repeat the exposure.