Operators Manual
4.3 Display Will Not Light
|
|
| Go to | |
|
|
| line: | |
1. | Is the instrument connected to line voltage? | YES | 2 | |
NO | 4 | |||
|
| |||
2. | Disconnect the line cord. Check the fuse. (See Section 4.6.) Is the fuse | YES | 3 | |
filament intact? | NO | 11 | ||
| ||||
3. | Attach known operational appliance to outlet. Does the appliance work? | YES | 4 | |
NO | 12 | |||
|
| |||
4. | Connect instrument to AC line. Press the Power button. | YES | 5 | |
Did the display light up? | NO | 7 | ||
| ||||
5. | Is the battery current between +0.10 A and +3.00 A? | YES | 6 | |
NO | 13 | |||
|
| |||
| The fault has been cleared. The battery is charging. Leave the instrument |
|
| |
6. | plugged in for at least 2 hours. When the battery is less than 5.4 volts and the |
|
| |
instrument is running on battery, the instrument will turn itself off. This is to |
|
| ||
|
|
| ||
| protect the battery from becoming severely discharged. |
|
| |
7. | Replace cord with known good cord or establish that the line cord is good. Is | YES | 8 | |
line cord good? | NO | 14 | ||
| ||||
8. | Was the battery low the last time the instrument was stored? | YES | 9 | |
NO | 16 | |||
|
| |||
9. | Has the instrument been plugged in for more than 24 hours? | YES | 10 | |
NO | 15 | |||
|
| |||
10. | PROBLEM: Instrument will not power up after 24 hours. See Section 4.10. |
|
| |
|
|
|
| |
11. | See Section 4.6. |
|
| |
|
|
|
| |
12. | Faulty Outlet. Use a different outlet. |
|
| |
|
|
|
| |
13. | See Section 4.2 to isolate problem. |
|
| |
|
|
|
| |
14. | ORDER: Proper line cord. See Section 4.9. |
|
| |
|
|
|
| |
| The battery may take more than 18 hours to recharge if the battery was in the |
|
| |
15. | low battery condition and was not recharged before storage. Leave the |
|
| |
| instrument plugged in and try again in |
|
| |
16. | PROBLEM: Instrument will not power up. See Section 4.10. |
|
| |
|
|
|
|