Troubleshooting and Service
Troubleshooting 4
| Example: 299.9 to 300.0 |
Cause: | Rounding error in measurement. |
Action: | None. |
Explanation: | The bias voltage is not actually changing. The control for the bias is in discrete steps. |
| The measurement is also in discrete steps. The control program may select a setting |
| that causes the measurement to be very near the transition between one measurement |
| step and the next. |