Operation Instructions B
the instrument will cease updating the exposure measurements when the current drops below the stop threshold and will return to measuring leakage.
NOTE
The instrument uses the start and stop threshold settings to sense when an exposure is in progress.
With the automatic leakage compensation settings displayed (Figure
button. The "" cursor will appear to the left of the ALC annunciator. Use the UP and DOWN arrow buttons to toggle the setting between on (ALC) and off (OFF).
With automatic leakage compensation enabled, press the TEST FUNCTION button to advance the "" cursor to the next field or exit Edit mode. When the cursor points to a threshold setting, use the UP and DOWN arrow buttons to increment and decrement the value, respectively. Nominal values for the start and stop thresholds are 1.3 and 1.2 pA, respectively. However, depending on the system leakage, other values may be necessary.
When automatic leakage compensation is disabled, the annunciator displays off (OFF) (Figure
ADC |
| 1.0 | ALC |
Start Thres: | pA | ||
Stop | Thres: | 0.9 | pA |
Current: | 0.0 | pA | |
measurements. |
|
| |
Figure | Automatic Leakage Compensation Enabled | ||
Figure | Automatic Leakage Compensation Disabled |
NOTE
The instrument may reset all measurements when
ADC | Thres: |
| OFF |
Start | NA pA | ||
Stop | Thres: | NA | pA |
Current: | 0.0 | pA |
the automatic leakage compensation parameters are changed.
NOTE
Disabling automatic leakage compensation prevents automatic exposure detection and the measurement of leakage current, and leakage measurements will display Not Available (NA).