B.6 Troubleshooting

This section outlines steps that the operator may take to identify and correct problems.

Symptom: "Configuration Error" displayed.

ADC

Configuration Error

Service Required

Error Number : 16

Cause: The instrument has lost the calibration or customization data.

Action: Convert the configuration error number to a binary value using a program such as the Windows® Calculator and use the table below to determine the extent of the problem. For example, a configuration error of “18” corresponds to a binary value of “10010” (bits 5 and 2 set), which indicates that the customization for both the ion chamber and settings are invalid.

Bit

Description

Action

1

Operator settings are

Select the proper settings

 

invalid.

from the front panel.

2

Customization settings are

Reinstall settings using the

 

invalid.

Customization software.

3

Electrometer calibration

Do NOT make

 

settings are invalid.

measurements. Return

 

 

instrument for repair.

4

Bias calibration settings are

Do NOT make

 

invalid.

measurements. Return

 

 

instrument for repair.

5

Ion chamber customization

Reinstall ion chambers using

 

settings are invalid.

the Customization software.

6

Ion chambers cannot be

Return instrument for repair

 

read.

to correct problem.

7

Additional operator settings

Select the proper settings

 

are invalid.

from the front panel.

8

Timer interval customization

Reinstall settings using the

 

settings are invalid.

Customization software.

9

Additional calibration

Do NOT make

 

settings are invalid.

measurements. Return

 

 

instrument for repair.

Explanation: The instrument detects whether the stored parameters are valid at power-up. If the parameters are invalid, it will affect one or more of the measurements. The operator may restore some of the information either by selecting options from the front panel or by running the customization software. Invalid calibration settings require a new calibration. Follow the recommendations in the above table.

Symptom: "Current Overrange" displayed.

B-12

Image 88
Contents 35040 Therapy DosimeterFluke Biomedical Table of Contents Addendum Features of Firmware Version 27JAN00 or Later Troubleshooting and ServiceSample Customization Reports Iii Blank Introduction FeaturesIntroduction Safety Symbols And Terms Before You Start Safety PrecautionsOperators Manual Specifications Factory Calibration Error Source ± %rdgMaximum Error Manual AddendaFull Scale Range Display Resolution Noise ± counts Full Scale Range Display ResolutionTimed Charge Measurements Current MeasurementsEffective Exposure Time Measurements SpecificationsGeneral Measurements High Performance Power Supply and Battery Charger Unpacking and Inspection 35040 Hardware Block Diagram of the Model 35040 Therapy Dosimeter Functional DescriptionPower Supply Front PanelMicroController Serial Communications PortBasic Operating Equations ElectrometerBias Supply Multiplexer and A/D ConverterQuantity Description Units Comments QExposureQAccum QAccumpreviousLeakage Measurements Charge and Current Measurements= time of electrometer conversion s Exposure Charge and Dose Timed Charge and DoseSetting Thresholds Predicted Timed Charge and DoseAccumulated Charge and Dose Exposure Current and RateExposure Time DoseRate = I a × Iccf Units/C × Adcf × RateMultiplier s⎝ P Air Density Correction FactorMm Hg = inch Hg x Standard Resolution Error Units Display Minimum MaximumSetup SetupFront Panel Familiarization IntroductionPowering The System Rear Panel FamiliarizationPower-Up Self-Test and Display Messages Ion Chamber ConnectorGeneral Considerations Basic Measurement TechniquesMaking Non-Timed Exposure Measurements Timer Use Timer Setup3Making Timed Exposure Measurements Battery Care Using Other Ion ChambersPower Battery CareOperation Instructions Front Panel ControlsPowering Up the Model 35040 Therapy Dosimeter Using the Test Function Button Powering Down the Model 35040 Therapy DosimeterOperation Instructions QScaleFactor=1.0000 IScaleFactor=1.0000 12. The Test Function Screen Showing the Timer Settings 5Automatic Leakage CompensationCalibrated by Keithley RMD Nov 15 6Timer Setup 17. The Test Function Screen Showing the Timer SettingsUsing the Detector Select Button Using the AIR Density ButtonUsing the Units Select Button Using the Bias Select ButtonSet 300.0 V 100% Central Electrde Pos IC Bias 300.0 V CEP Using the RESET/MEASURE Button Overranging ConditionsOverrange Conditions Reset ErrorCondition Values Overranged Reset Error With Automatic Leakage Compensation Enabled AnnunciatorsAnnunciators Annunciator ConditionMeasurement Indicators and Conditions Indicator Measurement Condition? left of value Condition Overranged MeasurementInstalling/Uninstalling Customization Software CustomizationConnection to the Computer Starting the Customization SoftwareMenu Items and Controls DownloadingCOM Port Drop Down List Close or File Close Ion ChambersFile Save AS and File Print Edit OK and Cancel Special ItemsDownload and Cancel 36. The Special Items Menu Calibration Temperature has focusMeasurement Displays Operators Manual Available Display Lines Display LineUsing the APPEND, DELETE, and Insert Buttons Display Line Group SelectionSelecting an Item and Placing It Changing a Line on a DisplayTimer Intervals 38. The Timer Intervals FormBias Settings 39. The Bias Settings FormExamples Changing the Pressure Units Customizing a Measurement DisplayAppending a Display Deleting a DisplayErrors Troubleshooting and Service TroubleshootingSymptom Display does not light up when Power is pressed Symptom ??Bias annunciator displayedSymptom HiBias annunciator displayed Symptom LoBias annunciator displayedSymptom Internal Reset Error displayed Symptom Configuration Error displayedBit Description Action Do not makeSymptom Current Overrange Displayed Symptom Dashes displayed instead of digitsSymptom Bias voltage measurement changes by one digit Measurements. ReturnExample 299.9 to Display Will Not Light Low Battery Hours? Is the battery voltage greater than 5.9 volts?Go to Bias ErrorLine Line Fuse Replacement Model 35040 Therapy Dosimeter Fuse ReplacementCleaning The Instrument Interface Connector CleaningReplaceable Parts Information Warranty Information Replaceable Parts ListRecalibration, Repair Information & Technical Assistance Replaceable Parts InformationBlank Sample Customization Reports Sample Documents35040 Appendix Sample Customization Reports 35040 Appendix Sample Customization Reports Blank Overview Introduction New FeaturesTimed Measurements Specifications Update Setup New Features BTimer Use Procedure Timer Setup ProcedureMaking Timed Exposure Measurements RESET/MEASURE Setup BADC ALC Automatic Leakage CompensationStart NA pA Stop Thres Current Operation Instructions BTimer Setup Decimal PlacesTable B-2. Measurement Indicators and Conditions Annunciators, Indicators, and Overrange ConditionsTable B-1. Annunciators Table B-3. Overrange Conditions Condition Overranged Measurement1Theory Of Operation Theory of OperationConfiguration Error Service Required Error Number Measurement Menu Main MenuMeasurement Displays Figure B-9 shows the Measurement Displays dialog box Table B-4. Available Display Lines Figure B-10. The Timer Intervals dialog box Page Fluke Biomedical