Operators Manual
4.5 Bias Error
|
|
| Go to | |
|
|
| line: | |
1. | Is "??Bias" displayed? | YES | 2 | |
NO | 6 | |||
|
| |||
2. | Has the instrument been on for more than one minute? | YES | 3 | |
NO | 8 | |||
|
| |||
3. | Turn off the power. Wait 10 seconds. Turn on the power. Wait 1 minute. Is | YES | 4 | |
"??Bias" displayed? | NO | |||
| ||||
4. | PROBLEM: Electrometer fault. See Section 4.10. | YES | 5 | |
NO | ||||
|
| |||
5. | Change the bias voltage to a different setting and then change back. Wait |
|
| |
about 10 seconds. Is the error on the display? |
|
| ||
|
|
| ||
| Change the bias voltage to 0.0 V. Disconnect all triax cables. Cap the | YES | 7 | |
6. | connectors. Change bias voltage to desired setting. Wait about 10 seconds. | |||
NO | 8 | |||
| Is the error on the display? | |||
|
|
| ||
7. | Change the bias voltage to 0.0 V. Examine the triax connectors on the |
|
| |
instrument, both front and rear, for foreign matter. Is foreign matter present? |
|
| ||
|
|
| ||
8. | See Section 4.7. | YES | 9 | |
NO | 18 | |||
|
| |||
9. | Normal operation established. | YES | 10 | |
NO | 11 | |||
|
| |||
10. | Do you use a separate cable to connect the ion chamber to the instrument? |
|
| |
|
|
|
| |
| Change the bias voltage to 0.0 V. Connect only the cable to the instrument. | YES | 12 | |
11. | Change the bias voltage to the desired setting. Wait about 10 seconds. Is the | |||
NO | 13 | |||
| error on the display? | |||
|
|
| ||
12. | PROBLEM: Bias fault in cable. See Section 4.10. |
|
| |
|
|
|
| |
| Change the bias voltage to 0.0 V. Connect the ion chamber. Change the bias | YES | 14 | |
13. | voltage to the desired setting. Wait about 10 seconds. Is the error on the | |||
NO | 16 | |||
| display? | |||
|
|
| ||
14. | PROBLEM: Bias fault in instrument. See Section 4.10. | YES | 15 | |
NO | 17 | |||
|
| |||
15. | PROBLEM: Bias fault in ion chamber. See Section 4.10. |
|
| |
|
|
|
| |
16. | This message is displayed until the signals from the isolated side are valid. |
|
| |
|
|
|
|