Fluke How to Use Electrical Safety Analyzer and Perform ESA612 Leakage Tests

Page 39

 

 

Electrical Safety Analyzer

 

 

Performing Electrical Safety Tests

 

Performing a Current Consumption Test

Which leakage tests are available depends on which

 

To measure the current consumed by the DUT, press

standard is selected. See the Selecting the Test Standard

 

section earlier in this manual to change the standard the

 

A. The Analyzer displays the current flowing through

 

Analyzer is using.

 

the mains connections of the test receptacle.

 

Table 5 lists six leakage current tests that have different

 

Performing Leakage Current Tests

 

names based on which standard is selected.

 

The Analyzer measures leakage current for a number of

 

Press Bto access the leakage current main menu

 

different DUT configurations. In addition to the leakage

 

shown in Figure 17.

 

found on the enclosure and the earth connection, the

 

 

 

Analyzer can measure leakage on each connected applied

 

 

part and combinations of connected applied parts.

 

 

Table 5. Test Names Based on Selected Standard

 

IEC60601

AAMI/NFPA 99

 

 

 

 

Protective Earth Resistance

Ground Wire Resistance

 

 

 

 

Earth Leakage Current

Ground Wire Leakage Current

 

 

 

 

Touch or Enclosure Leakage Current

Chassis Leakage Current

 

 

 

 

Patient Leakage Current

Lead to Ground Leakage Current

 

 

 

 

Patient Auxiliary Leakage Current

Lead to Lead Leakage Current

 

 

 

 

Mains on Applied Part (MAP) Leakage Current

Isolation Leakage Current

 

 

 

25

Image 39
Contents ESA612 Warranty and Product Support Contents Standard Terms and Conditions Certification Restrictions and Liabilities Table of Contents ESA612 Iii ESA612 List of Tables ESA612 List of Figures Viii Introduction Electrical Safety AnalyzerSymbol Description Safety InformationIntended Use Electrical Safety AnalyzerInstrument Familiarization Unpacking the AnalyzerInstrument Familiarization Side and Top-Panel Connections Side and Top-Panel Connections Name Connecting a DUT to the Analyzer Connecting to Line Power Turning the Analyzer OnTurning the Analyzer On Leakage Current Menu Accessing the Analyzer’s FunctionsSetting Up the Analyzer Setting Polarity Switching DelaySetting the Display Contrast Viewing Memory Performing Electrical Safety TestsSetting the Gfci Limit Setting up the BeeperSetting the Test Standard Performing Mains Voltage TestingPerforming a Ground Wire Protective Earth Resistance Test DUT is powered off for this test Electrical Safety Analyzer Fis112.eps Schematic Abbreviations Device Under Test Insulation Resistance Measurement Performing an Insulation Resistance TestFaw17.eps Applied Parts to Protective-Earth Insulation Test Schematic Mains to Applied Parts Insulation Test Schematic Mains to Non-Earth Accessible Conductive Points Schematic Applied Parts to Non-Earth Conductive Points Schematic Performing Leakage Current Tests Performing a Current Consumption TestMeasuring Earth Leakage Current Electrical Safety Analyzer Earth Leakage Current Test Schematic Performing a Chassis Enclosure Leakage Test Enclosure Leakage Current Test Schematic Performing a Lead-to-Ground Patient Leakage Lead-to-Ground Patient Leakage Current Test Schematic Tests Electrical Safety AnalyzerFaw30.eps Electrical Safety Analyzer Lead Isolation Mains On Applied Parts Leakage Test Schematic Performing an Alternative Applied Part Leakage Test Performing Electrical Safety TestsPerforming an Alternative Equipment Leakage TestAlternative Equipment Leakage Current Test Schematic Performing a Direct Equipment Leakage Test Alternative Applied Part Leakage Test Schematic Performing a Direct Applied Part Leakage Test When the EN62353 standard is selectedDirect Equipment Leakage Test Schematic Direct Applied Parts Leakage Current Test Schematic Performing a Differential Leakage Current Test Differential Leakage Current Test Schematic Using the 1210 Adapter Using the 1210 Adapter ESA612 EGC Lead Connection with 1210 Adapter Making Point-To-Point Measurements Measuring ResistanceMeasuring Voltage Measuring Current Simulating ECG WaveformsESA612 ECG Monitor Connections Storing Data into Memory Using MemoryControlling the Analyzer Remotely Viewing Memory DataDeleting Data from Memory Testing and Replacing the Fuses MaintenanceCleaning the Analyzer Cleaning the AnalyzerReplaceable Parts Fluke Biomedical Part Number Replaceable PartsReplaceable Parts Accessories Temperature SpecificationsVoltage Detailed SpecificationsDetailed Specifications ECG Performance Waveforms