Fluke ESA612 specifications Differential Leakage Current Test Schematic

Page 59

Electrical Safety Analyzer

Performing Electrical Safety Tests

 

MAINS

DEVICE UNDER TEST

 

ON

DUT_L1

L1

 

APPLIED

 

MD

 

PART

MAINS

L2

 

 

OPEN

DUT_L2

L2

 

 

 

REV

 

 

POL

 

 

 

FE

 

EARTH

 

 

OPEN

 

 

 

DUT_PE

 

PE

 

 

 

CONDUCTIVE PART

 

 

TEST

 

 

LEAD

 

 

faw32.eps

Figure 28. Differential Leakage Current Test Schematic

45

Image 59
Contents ESA612 Warranty and Product Support Contents Standard Terms and Conditions Certification Restrictions and Liabilities Table of Contents ESA612 Iii ESA612 List of Tables ESA612 List of Figures Viii Introduction Electrical Safety AnalyzerSymbol Description Safety InformationIntended Use Electrical Safety AnalyzerInstrument Familiarization Unpacking the AnalyzerInstrument Familiarization Side and Top-Panel Connections Side and Top-Panel Connections Name Connecting a DUT to the Analyzer Connecting to Line Power Turning the Analyzer OnTurning the Analyzer On Leakage Current Menu Accessing the Analyzer’s FunctionsSetting the Display Contrast Setting Up the AnalyzerSetting Polarity Switching Delay Viewing Memory Performing Electrical Safety TestsSetting the Gfci Limit Setting up the BeeperPerforming a Ground Wire Protective Earth Resistance Test Setting the Test StandardPerforming Mains Voltage Testing DUT is powered off for this test Electrical Safety Analyzer Fis112.eps Schematic Abbreviations Device Under Test Insulation Resistance Measurement Performing an Insulation Resistance TestFaw17.eps Applied Parts to Protective-Earth Insulation Test Schematic Mains to Applied Parts Insulation Test Schematic Mains to Non-Earth Accessible Conductive Points Schematic Applied Parts to Non-Earth Conductive Points Schematic Performing Leakage Current Tests Performing a Current Consumption TestMeasuring Earth Leakage Current Electrical Safety Analyzer Earth Leakage Current Test Schematic Performing a Chassis Enclosure Leakage Test Enclosure Leakage Current Test Schematic Performing a Lead-to-Ground Patient Leakage Lead-to-Ground Patient Leakage Current Test Schematic Tests Electrical Safety AnalyzerFaw30.eps Electrical Safety Analyzer Lead Isolation Mains On Applied Parts Leakage Test Schematic Performing an Alternative Applied Part Leakage Test Performing Electrical Safety TestsPerforming an Alternative Equipment Leakage TestAlternative Equipment Leakage Current Test Schematic Performing a Direct Equipment Leakage Test Alternative Applied Part Leakage Test Schematic Performing a Direct Applied Part Leakage Test When the EN62353 standard is selectedDirect Equipment Leakage Test Schematic Direct Applied Parts Leakage Current Test Schematic Performing a Differential Leakage Current Test Differential Leakage Current Test Schematic Using the 1210 Adapter Using the 1210 Adapter ESA612 EGC Lead Connection with 1210 Adapter Measuring Voltage Making Point-To-Point MeasurementsMeasuring Resistance Measuring Current Simulating ECG WaveformsESA612 ECG Monitor Connections Storing Data into Memory Using MemoryDeleting Data from Memory Controlling the Analyzer RemotelyViewing Memory Data Testing and Replacing the Fuses MaintenanceCleaning the Analyzer Cleaning the AnalyzerReplaceable Parts Fluke Biomedical Part Number Replaceable PartsReplaceable Parts Accessories Temperature SpecificationsVoltage Detailed SpecificationsDetailed Specifications ECG Performance Waveforms