ESA612
Users Manual
During the
Accessing the Analyzer’s Functions
For each test and setup function, the Analyzer uses a series of menus to access various Analyzer test and setup variables. As shown in Figure 5, the Analyzer indicates various leakage current tests along the bottom of the display. An Exit selection is also indicated as a way of backing out of the leakage current tests. Pressing a softkey (F1 through F4) under a specific test will cause the analyzer to setup for or perform the selected test.
In addition to the function softkeys, the Analyzer test functions may require using the navigation buttons to select parameters as well. In the example above, the leakage selection has Knext to it. This icon indicates the selection is controlled by pressing Gor H. In this example, the leakage current measurement is switched between AC+DC, AC only, or DC only. The applied parts indicator has Won the left end and Xon the right end.
These icons indicate the use of Eand Fto select an applied part.
fis102.jpg
Figure 5. Leakage Current Menu
The three buttons along the right side of the display
(PND) control the wiring of the Analyzer’s test receptacle for some electrical tests. The present state of these three buttons is displayed along the right edge of the display whenever these controls are active.
Figure 5 shows polarity is settable between normal, reversed, and off. Neutral is also settable to closed or open. Earth condition is not displayed, which indicates it can not be changed. However, earth is internally opened during this test.
10