| Electrical Safety Analyzer |
| Detailed Specifications |
Ranges | 0.0 to 199.9 μA |
| 200 to 1999 μA |
| 2.00 to 10.00 mA |
Accuracy |
|
DC to 1 kHz | ±(1 % of reading + (1 μA or 1 LSD, whichever is greater)) |
1 to 100 kHz | ±(2 % of reading + (1 μA or 1 LSD, whichever is greater)) |
100 kHz to 1 MHz | ±(5 % of reading + (1 μA or 1 LSD, whichever is greater)) |
Mains on applied part test voltage | 100 % ±5 % of Mains for AAMI, current limited to 1 mA ±25 % per AAMI |
| 100 % ±5 % of Mains for 62353 current limited to 3.5 mA ±25 % per 62353 |
| 100 % ±5 % of Mains for 60601 current limited to 7.5 mA ±25 % per 60601 |
Differential leakage |
|
Ranges | 10 to 199 μA |
| 200 to 2000 μA |
| 2.00 to 20.00 mA |
Accuracy | ±(10 % of reading + (2 counts or 20 μA, whichever is greater)) |
Insulation resistance |
|
Ranges | 0.5 to 20.0 MΩ |
| 20.0 to 100.0 MΩ |
Accuracy |
|
20 MΩ Range | ±(2 % of reading + 0.2 MΩ) |
100 MΩ Range | ±(7.5 % of reading + 0.2 MΩ) |
Source test voltage | 500 V dc (+20 %, |
Maximum load capacitance | 1 μF |
63