Fluke ESA612 Performing Electrical Safety Tests, Performing an Alternative Equipment Leakage

Page 51

 

Electrical Safety Analyzer

 

Performing Electrical Safety Tests

Performing an Alternative Equipment Leakage

The following outlet conditions apply when performing this

Test

test:

Note

The alternative equipment leakage test is available when the EN62353 standard is selected.

During the Alternative Equipment Leakage test, the voltage source is applied between short-circuited equipment outlet mains live, neutral, and equipment outlet earth, the exposed conductive surface on the housing, and all applied parts short-circuited together. Equipment is separated from mains during the test. The current which flows over the insulation of the DUT is measured.

This test is not applicable for equipment with internal electrical power source. The switches in mains part shall be closed during measurement.

To perform an alternative equipment leakage test:

1.Press B.

2.Press the softkey labeled Alternative Equipment.

3.Press Tto apply the voltage and read the current in the display.

Figure 24 shows the electrical connections between the Analyzer and the DUT during an Alternative Equipment Leakage Test.

Closed Earth

Open Earth

Note

If there are more than five applied parts to connect to the Analyzer, see Using the 1210 Adapter later in this manual.

Performing an Alternative Applied Part Leakage Test

Note

The Alternative applied part leakage test is available when the EN62353 standard is selected.

During the Alternative Applied Part Leakage test, the test voltage is applied between short-circuited applied parts of

asingle function and the short-circuited equipment outlet mains live, neutral, equipment outlet earth, and exposed conductive surface on the housing. This test should only be done for equipment with F-Type applied parts. For equipment with multiple applied parts, test each group of applied parts of a single function in turn with all others floating during the test. All applied parts can be connected to the Analyzer’s applied parts jacks and the lead selection will float those not selected.

37

Image 51
Contents ESA612 Warranty and Product Support Contents Standard Terms and Conditions Certification Restrictions and Liabilities Table of Contents ESA612 Iii ESA612 List of Tables ESA612 List of Figures Viii Introduction Electrical Safety AnalyzerSymbol Description Safety InformationIntended Use Electrical Safety AnalyzerInstrument Familiarization Unpacking the AnalyzerInstrument Familiarization Side and Top-Panel Connections Side and Top-Panel Connections Name Connecting a DUT to the Analyzer Connecting to Line Power Turning the Analyzer OnTurning the Analyzer On Leakage Current Menu Accessing the Analyzer’s FunctionsSetting Up the Analyzer Setting Polarity Switching DelaySetting the Display Contrast Viewing Memory Performing Electrical Safety TestsSetting the Gfci Limit Setting up the BeeperSetting the Test Standard Performing Mains Voltage TestingPerforming a Ground Wire Protective Earth Resistance Test DUT is powered off for this test Electrical Safety Analyzer Fis112.eps Schematic Abbreviations Device Under Test Insulation Resistance Measurement Performing an Insulation Resistance TestFaw17.eps Applied Parts to Protective-Earth Insulation Test Schematic Mains to Applied Parts Insulation Test Schematic Mains to Non-Earth Accessible Conductive Points Schematic Applied Parts to Non-Earth Conductive Points Schematic Performing Leakage Current Tests Performing a Current Consumption TestMeasuring Earth Leakage Current Electrical Safety Analyzer Earth Leakage Current Test Schematic Performing a Chassis Enclosure Leakage Test Enclosure Leakage Current Test Schematic Performing a Lead-to-Ground Patient Leakage Lead-to-Ground Patient Leakage Current Test Schematic Tests Electrical Safety AnalyzerFaw30.eps Electrical Safety Analyzer Lead Isolation Mains On Applied Parts Leakage Test Schematic Performing an Alternative Applied Part Leakage Test Performing Electrical Safety TestsPerforming an Alternative Equipment Leakage TestAlternative Equipment Leakage Current Test Schematic Performing a Direct Equipment Leakage Test Alternative Applied Part Leakage Test Schematic Performing a Direct Applied Part Leakage Test When the EN62353 standard is selectedDirect Equipment Leakage Test Schematic Direct Applied Parts Leakage Current Test Schematic Performing a Differential Leakage Current Test Differential Leakage Current Test Schematic Using the 1210 Adapter Using the 1210 Adapter ESA612 EGC Lead Connection with 1210 Adapter Making Point-To-Point Measurements Measuring ResistanceMeasuring Voltage Measuring Current Simulating ECG WaveformsESA612 ECG Monitor Connections Storing Data into Memory Using MemoryControlling the Analyzer Remotely Viewing Memory DataDeleting Data from Memory Testing and Replacing the Fuses MaintenanceCleaning the Analyzer Cleaning the AnalyzerReplaceable Parts Fluke Biomedical Part Number Replaceable PartsReplaceable Parts Accessories Temperature SpecificationsVoltage Detailed SpecificationsDetailed Specifications ECG Performance Waveforms