| Electrical Safety Analyzer |
| Performing Electrical Safety Tests |
Performing | different combinations of posts tied together: RA/R and |
Tests | LL/F, RA/R and LA/L, or LL/F and LA/L. |
Note
The
To measure the leakage current through each applied part or lead and selected combination of lead connections (all other or between two), press the softkey labeled Lead to Lead from the Leakage Test main menu shown in Figure 17. Figure 22 shows the electrical connections between the Analyzer and the DUT during a
The
After each post is isolated individually, the
fis107.eps
Figure 21. Applied Parts Connection Posts Display
Within the
Pswitches the polarity of the mains voltage applied to the Analyzer’s test receptacle between Normal, Off, Reverse, and Off. Pressing Nopens and closes the neutral connection to the Analyzer’s test receptacle. Pressing Dopens and closes the earth or ground connection to the Analyzer’s test receptacle.
Note
If there are more than five applied parts to connect to the Analyzer, see Using the 1210 Adapter later in this manual.
33