Varec 4532 manual Introduction, Installation and operation overview, Software history

Page 11

4532

Introduction

 

 

1Introduction

1.1 Introduction

The 4532 Average Temperature Sensor and Converter (ATC) is a multi-element (Pt100) resistance device designed to measure the average product temperature in bulk liquid storage tanks. The measured value is converted into a HART® compatible output for use in temperature compensated volumetric calculations. The 4532 ATC contains up to 6 temperature elements, spaced at 2 or 3 meter intervals along the length of the probe. Power and local HART communications are provided over a 2-wire, intrinsically safe (i.s.) connection from compatible instruments, such as the Varec 6000 Series Servo Tank Gauge (STG), 7500 Series Radar Tank Gauges (RTG) and the 4590 Tank Side Monitor (TSM).

1.2Installation and operation overview

Step 1. Installing on tank top

The actual installation may require an installer to work in a hazardous area. Safety must be taken into account in order to avoid any harmful conditions.

The installation method depends on the type of 4532 ATC. Refer to Chapter 4, Installation.

Step 2. Wiring to the host instruments (4590 TSM)

Wiring material and conditions must be in accordance with intrinsically safe standards.

One end (normally on the host instrument side) of the shield twisted pair of cables must be grounded at the terminal connection.

Refer to Chapter 4, Installation.

Step 3. 4532 ATC initial setup

Set up device settings for the 4532 ATC and the local HART setting to the host instrument.

Step 4. Data flow from the 4532 ATC to the host instruments

Individual element temperature data: Individual row element temperature can be accessed regardless of liquid level information on the 4532 ATC's data matrix.

Average temperature data: The host instrument sends liquid level data on the HART line to the 4532 ATC. The 4532 ATC calculates both Gas / Liquid phase average temperature based on this given liquid level.

1.3

Software history

 

 

 

 

 

 

 

 

 

Software version /

Software changes

Documentation

 

 

Date

 

changes

 

 

V1.45/04.2006

Original software

 

 

 

 

 

 

Varec, Inc.

1

Image 11
Contents Installation Page Copyright Disclaimer of Warranties Safety Precaution Definitions Safety Precautions Average Temperature/Water Bottom Sensor and Converter Contents Operation and Description of Instrument Function Index Varec, Inc Contents Software history Installation and operation overviewIntroduction IntroductionHart System functional diagramSafety instructions Return Power supplyWiring GroundingDirect voltage Safety instructions Scope of delivery IdentificationDevice designation Nameplate CE marks, declaration of conformity Design, dimensions InstallationUnpacking Recommended procedure Incorrect procedure Flexible tubeSpecial conditions for safe use Installation InstructionsInstallation Mounting instruction MountingTop anchor method Mounting on a fixed roof tankTensioning wire mounting Thermo well application Thermo well methodFlexible tube is stabilized by an anchor weight Anchor weight methodAnchor weight mounting Mounting on a floating roof tank Thermo well method 10Guide ring and anchor weight application Guide ring and anchor weight methodMounting Terminal Connection 1 4532 ATC terminal Wiring3 4590 TSM i.s. terminal 2 6000 Series Servo Tank Gauge terminal4532 ATC Grounding Wiring As a standalone generic Hart instrument Device set up 4590 TSMOperating Local Hart connection As a Varec tank gauging instrumentPreparation of 6000 Series Servo Tank Gauge 4532 ATC specific parameter set up on the 4590 TSMDevice set up 6000 Series Servo Tank Gauge 2 4532 ATC configuration on 6000 Series Servo Tank GaugeGVH450~459 Temp No.1~10 GVH480 DiagnosticGVH447 Reference Zero GVH449 ReferenceOperating Device Data Temperature measurement Primary values VH00 ~ VH09Operation and Description of Instrument Function Hart Device designationElement Temperature 2 VH20 ~ VH29 Element Temperature 1 VH10 ~ VH15Advanced VH28 Lower Limit SpotVH29 Upper Limit VH46 Hysteresis Width Item type read and write Default 10mmVH49 Liquid Offset VH48 Gas OffsetWB primary and Advanced temp VH50 ~ VH59 WB Adjustment and Operation Power VH60 ~ VH69VH53 Element Point Item type select Default VH55 Element Volume Item type read and write Range 1 ~VH69 Ref Voltage Temperature Adjustment VH70 ~ VH79VH71 Zero Adjust VH72 Adjust SpanVH76 Resistance Adj VH73 TemperatureVH74 Position VH75 ResistanceVH79 Protect Code VH81 Temperature Unit VH80 Present ErrorError code VH85 Kind of Interval VH82 Element NumberVH83 No. of Preambles VH84 Distance UnitVH90 Device ID Number Item type read and write Default Device setting 2 VH90~VH99VH88 Short Error VH89 Open ErrorVH96 Software Version VH92 Error Dis. SelVH94 Polling Address VH95 Manufacture IDRepairs to Ex-approved devices MaintenanceExterior cleaning RepairsGVH ReplacementSystem error messages Troubleshooting1Spare Parts Spare partsAnchor weight Low profile AccessoriesAnchor weight Tall profile 3Wire hook dimension Wire hook, Top anchorTechnical data at a glance Technical dataTechnical data Process connection SUS304 Order StructureOrder Structure Declaration of Contamination Form Declaration of ContaminationDeclaration of Contamination Form Index Index 4532 Your official representative Your offical representative