7Operating
7.1Local HART connection
7.1.1As a Varec tank gauging instrument
The 4532 ATC has been developed and designated primarily to work with the Varec tank gauging host instruments 4590 TSM or 6000 Series Servo Tank Gauge.
Temperature information is transmitted on a two wire i.s HART loop to the host instrument.
Since both the 4590 TSM and 6000 Series Servo Tank Gauge have a
7.1.2As a standalone generic HART instrument
The 4532 ATC is a HART foundation registered intrinsically safe loop powered device. The 4532 ATC provides four basic types of data as standard and parameter information via HART protocol, command 3. Configuration to host communication can be performed by a HC (Hand Communicator) or the Varec ToF field service tool to set a specific HART address.
20:Measurement function
• 0: Converter only
•1: Temperature + converter
These four basic data are available as standard.
•Average liquid temperature
•Average gas phase temperature
•Level (entered liquid level at "VH02 measured distance")
•Device status
7.2Device set up: 4590 TSM
Connect the loop powered HART communication cable from the 4590 TSM (intrinsically safe side compartment) to the 4532 ATC.
Since the 4590 TSM has been designed to recognize the 4532 ATC as a specific Varec
HART instrument, set up is easy.
7.2.1HART scanner
After the physical cabling between the 4532 ATC and the 4590 TSM is complete, scan all connected loop powered HART devices by activating "HART SCAN" on the version 1.x 4590 TSM. The version 2.x 4590 TSM continually scans the HART bus and will automatically detect the 4532 ATC when it is connected.
Caution! Not all 4590 TSM have fully accessible compatibility to recognize the 4532 ATC. Consult with your Varec representative to cross check the software and hardware version of the 4590 TSM.