Agilent Technologies E2625A, Option 100 manual To run the Mask Test conditionally

Page 23

Getting Started and Using Masks

To run the Mask Test conditionally

To run the Mask Test conditionally

1Access the Mask Test Setup dialog box and make sure the mask test is enabled.

2Click the Run Until selection arrow to view the options.

a If Stop on Failure is disabled, you can stop the mask test based on the following conditions:

Forever

The mask test will run continuously. This is the default selection.

Number of Waveforms

The mask test will run until the number of waveforms specified has been acquired. The number of waveforms is from 1 to 1,000,000,000.

Time

The mask test will run until the amount of time specified has elapsed. The time is in seconds, from 0.1 minute to 1440.0 minutes. The default is 1.0 minute.

bIf Stop on Failure is enabled, the first time the waveform crosses the mask violation region, the mask test will stop. See the next topic.

Click and select how you want the mask test to run.

Checkmark if you want the mask test to stop on the first failure.

3Select the Run Until option you want to use for the mask test.

2-13

Image 23
Contents User’s Guide This Book Getting Started and Using MasksContents Contents-2 Page Adding Option 100 to an existing Infiniium Oscilloscope What not to do Recommended accessoriesSupplied items Channels Bandwidth Sample Rate Oscilloscope modelsTo access the Mask Test Setup dialog box Is enabled Overview of the Mask Test Setup dialog boxGetting Started and Using Masks Introduction to Mask Testing Getting Started and Using MasksMake a copy of the Mask Test disk How Infiniium does Mask TestingLoad Mask dialog box will resemble To install the Standard MasksCal Memory Protect To calibrate the oscilloscopeCalibrating the probe adapter To choose a proper probe for Mask TestingTo get your waveform on the oscilloscope display To enable the Mask Test To load a Mask To align a Mask to the waveform Locate the Use File Setup When Aligning optionTo start and stop the Mask Test Starts mask test in the Mask Test Setup dialog boxTo understand the Mask Test tab description area To run the Mask Test conditionally Analyzing a failure To stop the Mask Test on the first failureStandard Masks Standard Masks Pre-Compliance Sonet Masks ITU G.703 International Standard Masks 2MbITU75.msk 6312kbITU110DS2.msk 6312kbITU75DS2.msk 8MbITU.msk 34MbITU.msk 44736kbITUDS3.msk 140MbITU0.msk 140MbITU1.msk 140MbITU1INV.msk 155MbITU0.msk 155MbITU1.msk 155MbITU1INV.msk Ansi T1.102 North American Standard Masks DS1CANSI.msk DS2ANSI.msk DS3ANSI.msk DS4NAANSI.msk DS4NAANSIMAX.msk STS1ANSIEYE.msk STS1ANSIPULSE.msk STS3ANSI.msk STS3ANSIMAX.msk FCC Part 68.308 Options A, B, and C Masks FCC68OPTC.msk Pre-Compliance Sonet Masks OC12SONET.msk Ieee 802.3 Masks 100BASETXSTP 100BASETXUTP Safety Document Warranty