Graco Essential ProMix 2KE Parts Manual for Troubleshooting and Maintenance

Page 9

Troubleshooting

Isolate a Mixing Problem

A mixing problem can be caused by a problem with the controller, meters, and solenoid valves, as well as the mix manifold.

1.To isolate the problem, check for any visible faults or errors:

a.Are all air and fluid tubes, hoses, and electrical cables properly connected?

b.Are valves and controls properly set?

c.Do the fluid supplies, solenoids, and spray gun have sufficient air pressure?

d.Do the fluid supplies need refilling?

2.If there is a process control problem, refer to your controller manual.

Common Causes

The flow rate is too high.

Highly unbalanced pressures from the fluid supply system.

Slow actuation of component A or B dispense valves.

System leaks.

Troubleshooting

Unbalanced Pressure

1.Check all component fluid supply pressures.

2.If the fluid supply pressures are not about equal, adjust their fluid pressure regulators, until the pres- sures are about the same.

3.If the pressures are already about equal, verify that the dose valves are operating properly.

Dose Valve Operation

Manually operate the dose valves by actuating their solenoids. The valves should snap open and shut quickly. If the valves move slowly, it could be caused by:

Air pressure to the valve actuators is too low. Minimum: 75 psi (0.52 MPa, 5.2 bar) Recommended: 85 psi (0.6 MPa, 6.0 bar).

Valve actuating air constricted by dirt or water in the air supply.

Solenoid or tubing restricted.

Dose valve seals need lubrication (see manual 312782).

Air piston o-rings and packings are not lubricated (see manual 312782),

Valve setting is turned out too far. See Mix Manifold Valve Settings, page 8.

312781E

9

Image 9 Contents
Fluid Mix Manifold Important Safety InstructionsContents Related ManualsModels Flush Before Using Equipment InstallationAir Connections Fluid ConnectionsKey ProMix 3KS Fluid Station Key ProMix 2KS Fluid Station Key DVA2 DVA3 behind Restrictor Sizes Size Code Orifice Size Setup the Fluid Manifold for Dynamic DosingOperation Adjust gun atomizing air pressure as neededPurging Operation GuidelinesMix Manifold Valve Settings MaintenancePreventive Maintenance Schedule WeeklyIsolate a Mixing Problem TroubleshootingCommon Causes Unbalanced PressureRemove the Integrator Mixer RepairRemove the Restrictor ProMix 2KE, Dynamic Dosing only Remove from the bottomDispense Valve Cap Repair the Valves and Seats ProMix 2KS and ProMix 3KS onlyReplace the Outlet Tube and Static Mixer Tube Rebuild the Mix ManifoldRepair 312781E Part No , for ProMix 2KS Proportioner PartsIntegrator Kits include parts Part No. Description QtySize Description Part No , for ProMix 3KS Proportioner Part No , for ProMix 3KS Proportioner Part No , for ProMix 2KE Proportioner, Sequential Dosing Integrator KitsPart Description Qty Restrictor Kit Part No , for ProMix 2KE Proportioner, Dynamic DosingParts 289695, 256875, and 262398 4000 psi 28 MPa, 280 bar Technical DataGraco Information Graco Standard Warranty