Agilent Technologies 6051A, 6050A manual Calibration Flowcharts, Example Programs

Page 77

CALibration:SAVE

Writes the present calibration constants into the EEprom. This command does not have to be sent until all ranges and modes have been calibrated. If the unit is turned off before CAL:SAVE is sent, the new calibration constants are lost.

Calibration Flowcharts

The flowcharts in Figures 6-2 and 6-3 describe the calibration procedures for "A" and "B" modules, respectively. They correspond to the example calibration programs. The flowcharts indicate the appropriate statement that is used in the program example to accomplish each step. They also indicate when to set the power supply to the appropriate voltage and current output. Refer to the Calibration Information (Table 4) in your module manual for the variable values, power supply settings, and current shunts.

Calibration mode is turned on at the beginning of the calibration procedure. Remember to save the calibration constants after you have verified that they are within specifications. Do not turn calibration mode off until after you have saved the new calibration constants - otherwise the new calibration constants will be lost.

Note

When calibrating the high calibration point of the high current range and high current transient level, you

 

must wait about 30 seconds for the internal current shunt of the module to stabilize with the full current

 

applied before you execute the CAL:MEAS:HIGH command. Because the high current range

 

calibration cause the module to heat up, you should also allow about 30 seconds time for the module to

 

cool down to room temperature before continuing to calibrate any other modes or ranges.

One shortcut that is used in this calibration procedure is that the readback DAC is calibrated for current readback after the high current range calibration, and calibrated for voltage readback after the voltage range calibration. This is because the readback setups are the same as the setups for the high current and voltage ranges. Another shortcut is that the same values are used to calibrate the main DAC as well as the readback DAC. You may wish to use different values to calibrate the readback DAC to optimize accuracy.

It is not necessary to calibrate the current readback for the low current range or for reading back resistance values. This is because the high current readback calibration takes care of the low current range. The resistance values that are read back are calculated based on the voltage at the input terminals and the current through the internal current shunt resistor. If the readback DAC has been calibrated for voltage and current readback, resistance readback will be accurate.

Note

Remember to turn the unit off after you have saved the new calibration constants. When the unit is turned

 

on again, the new calibration constants are used to recalculate the software OP and OC limits.

 

 

Example Programs

Each example program in this chapter is written in the, Agilent BASIC Language. If you are using an HP Series 200/300 computer, simply type in the program and run it. If you are using a different computer or programming language, you will have to modify the program before you can run it.

The program can be used to calibrate all of the modules. You must specify the Electronic Load address and the channel number of the module that you are calibrating as shown in lines 10 and 20. (The program assumes address 705, channel 1.) Then you must make the variable assignments for the module that you are calibrating in lines 40 through 90. Refer

to the Calibration Information table of the module you are calibrating for the values to assign each variable. Do not change the last value (Flag) in lines 40, 50, 70, 80, and 90.

When the program is run, it will stop at appropriate places and prompt you to set the power supply according to the calibration table, enter your measured values into the computer, and verify that the values are within specifications.

Calibration 77

Image 77
Contents Operating Manual Certification Safety Summary Symbol Description Safety SummaryHerstellerbescheinigung Manufacturer’s DeclarationPrinting History Page Table of Contents Local Operation InstallationCalibration Considerations for Operating in Constant Resistance ModeRemote Operation Page General Information What’s In This ManualOptions Specifications Safety RequirementsDimensions Page Operation Overview IntroductionRemote Programming Local/Remote ControlFront Panel Description Modes of Operation Extended Power OperationProgrammable Features Triggered Current Level Constant Current CC ModeImmediate Current Level Transient Current Level Constant Resistance CR ModeSoftware Current Limit Slew RateImmediate Resistance Level Constant Voltage CV ModeTriggered Resistance Level Transient Resistance LevelTransient Operation Triggered Voltage LevelTransient Voltage Level Pulsed Transient Operation Hpsl Command DescriptionContinuous Transient Operation Sets pulse width to 1 millisecond Selects the external trigger inputSelects pulsed transient operation Selects the external trigger input source Triggering a preset levelSelects toggled operation Triggering a transient pulseRisetime Transition Limitation Slew Rate And Minimum Transition TimeTransition Times and Slew Rates Short On/Off Input Current, Voltage, and Power MeasurementInput On/Off Saving and Recalling SettingsReading Remote Programming Errors Status Reporting Protection FeaturesResetting Latched Protection Overcurrent OverpowerOvervoltage Overtemperature Control ConnectorReverse Voltage Remote SensingExternal Programming Input FaultPage Inspection Installing The ModulesPower Cord Configurations Procedure Channel Number Installing The MainframesRack Mounting Turn-On CheckoutCooling Changing Line Voltage Line Voltage SwitchesChannel Errors Description Gpib ErrorsTurn-On/Selftest Display DescriptionController Connection Power TestWire Size Strip back Rear Panel Connectors and SwitchesGpib Address AWGSense Switch Input Binding Post Control Connector+Sand -S Pins Al and A2IM and VM Com pin A3Application Connections Trigger ConnectorWiring Considerations Ampacity Stranded Copper Wire Ampere Capacity Wire SizeLocal Sense Connections Remote Sense ConnectionsZero-Volt Loading Connections Maximum Wire Lengths to Limit Voltage Drops12. Local Sensing 14. Parallel Operation Page Local Operation Local OperationControls and Indicators Description Chan Keys Function Keys Local Control Overview Using The Chan Keys Selecting the Channel Using The Function KeysIdentifying the Selected Channel Turning the Input On/OffRecommended Programming Sequence Setting CC Values Setting the Mode of OperationProgramming Ranges ExamplesSetting CR Values Examples Programming Range Setting CV ValuesTransient Operation Shorting The Input Displaying Error Codes Using The System KeysSetting The Gpib Address Recalling the Factory Default Values Changing Wake-up SettingsPage Gpib Address Enter/Output StatementsOutput EnterSelecting a Channel Sending a Remote CommandGetting Data Back Output 705 MeascurrRemote Programming Commands CV Mode Example CC Mode ExampleOutput 705INPUT on Output 705MEASCURR? Output 705 Chan 2INPUT OFF Output 705MODEVOLTRemote Programming Flowchart Sheet Remote Programming Flowchart Sheet Output 705INPUT on Output 705MEASPOW? CR Mode ExampleContinuous Transient Operation Example Output 705CHAN 2INPUT OFF Output 705MODECURROutput 705CHAN 1INPUT OFF Output 705MODEVOLT Pulsed Transient Operation ExampleSynchronous Toggled Transient Operation Example Output 705 Trigsour TIM Page Calibration Equipment RequiredCharacteristics Calibration CommandsEquipment Required for Calibration Recommended ModelExample Programs Calibration FlowchartsCalibration Flowchart for a Modules Calibration Flowchart for a Modules Calibration Flowchart for a Modules Pause Subend Program Listing for a ModulesPause Print Voltage Calibration Line ElseEND if 610 Set low calibration point Calibration Flowchart for B Modules Calibration Flowchart for B Modules Calibration Flowchart for B Modules Clear Screen Print TABXY10,10CALIBRATION Done Program Listing for B ModulesWait 1260 If Flag then 1270 Output @LdRESReshipt 1280 Considerations For Operating In Constant Resistance Mode Considerations For Operating In Constant Resistance ModeConsiderations For Operating In Constant Resistance Mode Index IndexIndex Index 19, 20, 21 Agilent Sales and Support Offices Agilent Sales and Support OfficesManual Updates

6051A, 6050A specifications

Agilent Technologies has long been a leader in providing high-performance test and measurement solutions, and the 6050A and 6051A models exemplify this commitment to quality and innovation. The 6050A and 6051A are versatile signal generators that cater to a diverse range of applications, including research and development, manufacturing, and education, making them essential tools in laboratories and production environments.

The Agilent 6050A is a high-performance RF signal generator known for its frequency range capabilities, which span from 100 kHz to 20 GHz. It offers exceptional phase noise performance and low harmonic distortion, making it ideal for applications that require high signal integrity. The device supports various modulation formats, including AM, FM, and pulse modulation, allowing users to generate a wide range of test signals to simulate real-world conditions.

The 6051A builds upon the robust features of the 6050A with enhanced specifications and additional functionalities. It features a larger frequency modulation bandwidth, pushing the envelope for applications requiring more complex signal generation. The 6051A showcases a superior output power range, ensuring that test signals can be reliably produced at varying power levels. This model also includes advanced output control options that allow for precise signal manipulation, making it particularly suited for testing amplifiers and other RF components.

Both models share core technologies that ensure reliable performance, such as direct digital synthesis (DDS) and phase-locked loop (PLL) architectures. These technologies contribute to the exceptional frequency stability and accuracy that engineers and scientists have come to rely on. Additionally, the user-friendly interface integrated into both models simplifies operation and allows for quick configuration changes, facilitating efficient research and testing workflows.

With comprehensive connectivity options, including GPIB, USB, and Ethernet, the 6050A and 6051A can easily integrate into automated test environments. Their reliability, performance, and flexibility make them a perfect choice for those looking to advance their testing capabilities, whether in academic research, product development, or quality assurance in manufacturing.

In summary, the Agilent Technologies 6050A and 6051A signal generators are powerful tools designed to meet the demands of modern RF testing. Their advanced features, paired with Agilent’s reputation for quality and precision, make them invaluable assets in any engineering or research portfolio. Whether you require sophisticated signal generation for prototype testing or educational purposes, these models will deliver the performance needed to support your objectives.