Teledyne 633 user manual Manually lifting/lowering chamber optical head

Page 38

Model 633 Aethalometer® Black Carbon Monitor User Manual

Teledyne Advanced Pollution Instrumentation

3.4 Manually lifting/lowering chamber optical head

In case the instrument is turned off, the chamber can be lifted manually, as follows:

(Stop measurements first: Operation>General>Stop).

Note the retention pin at the top of the chamber optical head.

Grasp the optical head securely in both hands with fingers at bottom and thumbs on top, pressing retention pin down and raising optical head up.

Figure 11. Manually Lifting the Chamber Optical Head

ATTENTION! – If the chamber is lifted while instrument is operating (STOP button not pressed), it is not possible to restart the measurement without cycling the instrument power. (Flip both power switches to OFF; then flip each to ON).

36

07645A DCN6170

Image 38
Contents Model Aethalometer Black Carbon Monitor Page Trademarks This page intentionally left blank Table of Contents Spare Parts List Introduction Description1 ‘DualSpot’ Technology Teledyne Advanced Pollution Instrumentation 07645A DCN6170 Teledyne Advanced Pollution Instrumentation Automatic Zero and Span User and Communications Interfaces Modular Construction Technical Specifications Model 633 functional block diagram Functional DescriptionEnclosure USB Connector rear panel AIR Inlet and Outlet ConnectorsFlowmeters ROLL­ON and UN­ROLL Tape SpoolsUSB Connectors front panel Display and TOUCH­SCREENON/OFF Switch This page intentionally left blank Safety Notes and Labels Fire and explosion Instrument service and repair Moving parts Unpacking the system Sampling lineConnecting the sampling line Powering on the Aethalometer Initialization screen with test results Filter tape installation 07645A DCN6170 Teledyne Advanced Pollution Instrumentation Tape direction is from bottom, left to right Teledyne Advanced Pollution Instrumentation 07645A DCN6170 Changing filter tape Manually lifting/lowering chamber optical head Manually Lifting the Chamber Optical HeadUser INTERFACE, Settings and Operation User interface and settings07645A DCN6170 Instrument Status Status Parameter Condition DescriptionDownloading and Viewing Data Header Data lineData field description Connecting to External Datalogger or PC Serial Commands for Communication with the Aethalometer External Devices Supported devicesThis page intentionally left blank Maintenance and Service FrequencyCleaning the cyclone Teledyne Advanced Pollution Instrumentation Leak test Report Automatic flow calibration procedure Automatic flow calibration procedure Automatic flow calibration procedure Manual flow calibration procedure Manual flow meter calibrationManual flow calibration procedure Manual flow calibration procedure Manual flow calibration procedure Manual flow calibration procedure Manual flow calibration procedure Manual clean air test Manual stability testCleaning the optical head Optical head cleaning procedure Optical head top Optical head platform Optical head bottom Teledyne Advanced Pollution Instrumentation 07645A DCN6170 Teledyne Advanced Pollution Instrumentation Calibrating the tape sensor Teledyne Advanced Pollution Instrumentation This page intentionally left blank Technical Support and Contact Information This page intentionally left blank Spare Parts List