Model 633 Aethalometer® Black Carbon Monitor User Manual

Teledyne Advanced Pollution Instrumentation

Data field description:

­Date(yyyy/MM/dd): date.

­Time(hh:mm:ss): time.

Timebase: timebase – units in seconds.

­Ref1, Sen1Ch1, Sen2Ch1…: are the raw signal Reference (Ref), Sense 1 ( Sen1) and Sense 2 ( Sen2) values form which the BC concentrations are calculated for channel 1 (Ch1), that is wavelength 370 nm. BC11 is the uncompensated BC calculated from the spot 1 for channel 1. BC1 is the final results the BC calculated form measurements for channel 1 (370 nm).

­Flow1; Flow2; FlowC: Measured flow in ml/min. Flow 1 is flow through the spot 1, Flow_C is common (total flow) through the optical chamber, Flow_2 is the difference between these two.

­Pressure (Pa); Temperature (°C); RH (%): These are the pressure and temperature which the instrument uses to report flow. By chosing to report mass flow, (101325 Pa, 21.11 C) are used to convert the measured mass flow to reported volumetric flow at these conditions (these are values used by the flow meter in the AE33). These values can be substituted for any other (pressure, Temp) values, should you want the flow to be reported at 0 C, or 25 C for example. Optionally, you can choose volumetric flow at ambient conditions to be reported. To do that you need an optional weather station accessory for the (pressure, Temp) measurement. We offer this separately (PN DU00000200). This external device is connected to the COM ports, please see below.

ContTemp; SupplyTemp: control and power supply board temperatures

­Status; ContStatus; DetectStatus; LedStatus; ValveStatus; LedTemp: status codes for the instrument, reported in the , please see manual, chapter 5.2. and the subcomponents: controller, detector LED board; valve status (each valve can be on or off); LED board temperature

­BC11; BC12; BC1; BC21; BC22; BC2; BC31; BC32; BC3; BC41; BC42; BC4; BC51; BC52; BC5; BC61; BC62; BC6; BC71; BC72; BC7: BC11 is the uncompensated BC calculated from the spot 1 for channel 1. BC1 is the final results the BC calculated form measurements for channel 1 (370 nm).

42

07645A DCN6170

Image 44
Contents Model Aethalometer Black Carbon Monitor Page Trademarks This page intentionally left blank Table of Contents Spare Parts List Introduction Description1 ‘DualSpot’ Technology Teledyne Advanced Pollution Instrumentation 07645A DCN6170 Teledyne Advanced Pollution Instrumentation Automatic Zero and Span User and Communications Interfaces Modular Construction Technical Specifications Model 633 functional block diagram Functional DescriptionEnclosure USB Connector rear panel AIR Inlet and Outlet ConnectorsFlowmeters ROLL­ON and UN­ROLL Tape SpoolsUSB Connectors front panel Display and TOUCH­SCREENON/OFF Switch This page intentionally left blank Safety Notes and Labels Fire and explosion Instrument service and repair Moving parts Unpacking the system Sampling lineConnecting the sampling line Powering on the Aethalometer Initialization screen with test results Filter tape installation 07645A DCN6170 Teledyne Advanced Pollution Instrumentation Tape direction is from bottom, left to right Teledyne Advanced Pollution Instrumentation 07645A DCN6170 Changing filter tape Manually lifting/lowering chamber optical head Manually Lifting the Chamber Optical HeadUser INTERFACE, Settings and Operation User interface and settings07645A DCN6170 Instrument Status Status Parameter Condition DescriptionDownloading and Viewing Data Header Data lineData field description Connecting to External Datalogger or PC Serial Commands for Communication with the Aethalometer External Devices Supported devicesThis page intentionally left blank Maintenance and Service FrequencyCleaning the cyclone Teledyne Advanced Pollution Instrumentation Leak test Report Automatic flow calibration procedure Automatic flow calibration procedure Automatic flow calibration procedure Manual flow calibration procedure Manual flow meter calibrationManual flow calibration procedure Manual flow calibration procedure Manual flow calibration procedure Manual flow calibration procedure Manual flow calibration procedure Manual clean air test Manual stability testCleaning the optical head Optical head cleaning procedure Optical head top Optical head platform Optical head bottom Teledyne Advanced Pollution Instrumentation 07645A DCN6170 Teledyne Advanced Pollution Instrumentation Calibrating the tape sensor Teledyne Advanced Pollution Instrumentation This page intentionally left blank Technical Support and Contact Information This page intentionally left blank Spare Parts List