5 MAINTENANCE and SERVICE

The Aethalometer needs to be maintained regularly. We recommend the following schedule for regular maintainance.

Item

Sample inlet flow

Inspect the sample line tubing

Inspect and clean the size selective inlet (if present)

Clean cyclone

(5.1 Cleaning the cyclone)

Verify time and date (if not set to update automatically) (Time & Date in OPERATION>GENERAL page).

Inspect optical chamber, clean if necessary

(Section 5.6 Cleaning the optical head)

Check flow, calibrate if necessary

(Section 5.2 Leak test).

(For calibration: Section 5.3 Automatic flow meter calibration or Section 5.4 Manual flow meter calibration).

Change by‐pass cartridge filter

(Section 5.7 Changing the bypass cartridge filter)

Install a new filter tape roll

(Section 3.4 Filter tape installation)

Calibrate tape sensor

(Section 5.8 Calibrating the tape sensor)

Frequency

Monthly

Monthly

Monthly

Weekly at first to gauge debris accumulation, then adjust as needed.

Monthly

Semi‐annually

Site dependent, use educated judgement!

Semi‐annually or as needed.

Annually or as needed.

As needed. The instrument issues a warning.

When measurement chamber has been removed/replaced.

07645A DCN6170

47

Image 49
Contents Model Aethalometer Black Carbon Monitor Page Trademarks This page intentionally left blank Table of Contents Spare Parts List Description Introduction1 ‘DualSpot’ Technology Teledyne Advanced Pollution Instrumentation 07645A DCN6170 Teledyne Advanced Pollution Instrumentation Automatic Zero and Span User and Communications Interfaces Modular Construction Technical Specifications Functional Description Model 633 functional block diagramEnclosure AIR Inlet and Outlet Connectors USB Connector rear panelROLL­ON and UN­ROLL Tape Spools FlowmetersDisplay and TOUCH­SCREEN USB Connectors front panelON/OFF Switch This page intentionally left blank Safety Notes and Labels Fire and explosion Instrument service and repair Moving parts Sampling line Unpacking the systemConnecting the sampling line Powering on the Aethalometer Initialization screen with test results Filter tape installation 07645A DCN6170 Teledyne Advanced Pollution Instrumentation Tape direction is from bottom, left to right Teledyne Advanced Pollution Instrumentation 07645A DCN6170 Changing filter tape Manually Lifting the Chamber Optical Head Manually lifting/lowering chamber optical headUser interface and settings User INTERFACE, Settings and Operation07645A DCN6170 Status Parameter Condition Description Instrument StatusDownloading and Viewing Data Data line HeaderData field description Connecting to External Datalogger or PC Serial Commands for Communication with the Aethalometer Supported devices External DevicesThis page intentionally left blank Frequency Maintenance and ServiceCleaning the cyclone Teledyne Advanced Pollution Instrumentation Leak test Report Automatic flow calibration procedure Automatic flow calibration procedure Automatic flow calibration procedure Manual flow meter calibration Manual flow calibration procedureManual flow calibration procedure Manual flow calibration procedure Manual flow calibration procedure Manual flow calibration procedure Manual flow calibration procedure Manual stability test Manual clean air testCleaning the optical head Optical head cleaning procedure Optical head top Optical head platform Optical head bottom Teledyne Advanced Pollution Instrumentation 07645A DCN6170 Teledyne Advanced Pollution Instrumentation Calibrating the tape sensor Teledyne Advanced Pollution Instrumentation This page intentionally left blank Technical Support and Contact Information This page intentionally left blank Spare Parts List