5.5 Manual stability test

Stability test is conducted without flow to determine the performance of the light source – detector pair. In the Operation/General tab select the desired timebase and click “Stability” button to start the test. Stability test is marked by status 1024. During the test BC values will be calculated assuming there is 5 LPM air flow. Manually stop the test when the desired test duration is complete.

Interpretation of the results:

-Average BC values should be close to zero if the Aethalometer is warmed up (was measuring for a few hours just before conducting the test).

‐ Point to point variation of BC (PPBC) is calculated as an average absolute difference between the consecutive BC measurements. It depends on the timebase, tested channel and spot. For example at 1 s timebase PPBC61 = approx. 400 ng/m3.

5.5Manual clean air test

Clean air test is conducted using the built‐in filter to determine the performance of Aethalometer under the standard flow condition. In the Operation/General tab select the desired timebase and flow, then click “Clean air” button to start the test. Clean air test is marked by status 2048. Manually stop the test when the desired test duration is complete.

Interpretation of the results:

Average BC values should be close to zero if the Aethalometer is warmed up (was measuring for a few hours just before conducting the test). It is possible to observe a short transient due to filter compression.

Point to point variation of BC (PPBC) is calculated as an average absolute difference between the consecutive BC measurements. It depends on the selected timebase, flow and tested channel. For example PPBC61 = approx. 500 ng/m3 at 5 LPM and 1 s timebase.

07645A DCN6170

61

Image 63
Contents Model Aethalometer Black Carbon Monitor Page Trademarks This page intentionally left blank Table of Contents Spare Parts List Description Introduction1 ‘DualSpot’ Technology Teledyne Advanced Pollution Instrumentation 07645A DCN6170 Teledyne Advanced Pollution Instrumentation Automatic Zero and Span User and Communications Interfaces Modular Construction Technical Specifications Functional Description Model 633 functional block diagramEnclosure AIR Inlet and Outlet Connectors USB Connector rear panelROLL­ON and UN­ROLL Tape Spools FlowmetersDisplay and TOUCH­SCREEN USB Connectors front panelON/OFF Switch This page intentionally left blank Safety Notes and Labels Fire and explosion Instrument service and repair Moving parts Sampling line Unpacking the systemConnecting the sampling line Powering on the Aethalometer Initialization screen with test results Filter tape installation 07645A DCN6170 Teledyne Advanced Pollution Instrumentation Tape direction is from bottom, left to right Teledyne Advanced Pollution Instrumentation 07645A DCN6170 Changing filter tape Manually Lifting the Chamber Optical Head Manually lifting/lowering chamber optical headUser interface and settings User INTERFACE, Settings and Operation07645A DCN6170 Status Parameter Condition Description Instrument StatusDownloading and Viewing Data Data line HeaderData field description Connecting to External Datalogger or PC Serial Commands for Communication with the Aethalometer Supported devices External DevicesThis page intentionally left blank Frequency Maintenance and ServiceCleaning the cyclone Teledyne Advanced Pollution Instrumentation Leak test Report Automatic flow calibration procedure Automatic flow calibration procedure Automatic flow calibration procedure Manual flow meter calibration Manual flow calibration procedureManual flow calibration procedure Manual flow calibration procedure Manual flow calibration procedure Manual flow calibration procedure Manual flow calibration procedure Manual stability test Manual clean air testCleaning the optical head Optical head cleaning procedure Optical head top Optical head platform Optical head bottom Teledyne Advanced Pollution Instrumentation 07645A DCN6170 Teledyne Advanced Pollution Instrumentation Calibrating the tape sensor Teledyne Advanced Pollution Instrumentation This page intentionally left blank Technical Support and Contact Information This page intentionally left blank Spare Parts List