Teledyne 633 user manual Connecting to External Datalogger or PC

Page 45

Teledyne Advanced Pollution Instrumentation

Model 633 Aethalometer® Black Carbon Monitor User Manual

­K1; K2; K3; K4; K5; K6; K7: K_i are the compensation parameters for wavelengths i=1…7

­TapeAdvCount: TapeAdvCount – tape advances since start

ID_com1; ID_com2; ID_com3; fields_i: ID_comi are the identifiers telling the Aethalometer which auxiliary device is connected to which serial port (necessary because of the different data structure). This is a 3 byte field:

0 2 0 21.1

means that the “Comet temperature probe” (instrument code 2) is connected to COM2. The temperature is 21.1 C. (This is not true for the temperature and pressure measurements to calculate volumetric flow using AMES_TPR159 sensor!)

If we connected additionally another aux instrument to COM1 the dataline would be (ID_comi:

5 2 0):

Date(yyyy/MM/dd); Time(hh:mm:ss); Timebase; RefCh1; Sen1Ch1; Sen2Ch1; RefCh2; Sen1Ch2; Sen2Ch2; RefCh3; Sen1Ch3; Sen2Ch3; RefCh4; Sen1Ch4; Sen2Ch4; RefCh5; Sen1Ch5; Sen2Ch5; RefCh6; Sen1Ch6; Sen2Ch6; RefCh7; Sen1Ch7; Sen2Ch7; Flow1; Flow2; FlowC; Pressure (Pa); Temperature (°C); RH (%); ContTemp; SupplyTemp; Status; ContStatus; DetectStatus; LedStatus; ValveStatus; LedTemp; BC11; BC12; BC1; BC21; BC22; BC2; BC31; BC32; BC3; BC41; BC42; BC4; BC51; BC52; BC5; BC61; BC62; BC6; BC71; BC72; BC7; K1; K2; K3; K4; K5; K6; K7; TapeAdvCount; ID_com1; ID_com2; ID_com3; fields_i

2012/09/21 00:34:00 60 890416 524323 709193 823296 573862 756304 884844 619592 789142 822391 673266 816066 792706 686925 828401 738101 718325 841075 789053 722690 833686 3325 1674 4999 101325 21.11 ‐1 30 40 0 0 10 10 00000 0 1150 1290 1242 1166 1248 1215 1150 1231 1190 1146 1196 1175 1214 1195 1234 1144 1114 1139 1180 1225 1174 0.00133 0.00095 0.00092 0.00080 0.00057 ‐0.00024 ‐0.00025 12 5 2 0 1090 45 20.0 21.1

where the aux instrument with the code 5 is a weather station reporting pressure (1090 mbar), RH (45%) and temperature (20.0). the temperature 21.1 is reported by the “Comet temperature probe” installed at a different location.

4.4 Connecting to External Datalogger or PC

Connect with a crossover cable (or standard serial cable with null modem adapter) from the PC / Datalogger to the serial port of the instrument on the rear panel.

07645A DCN6170

43

Image 45
Contents Model Aethalometer Black Carbon Monitor Page Trademarks This page intentionally left blank Table of Contents Spare Parts List Description Introduction1 ‘DualSpot’ Technology Teledyne Advanced Pollution Instrumentation 07645A DCN6170 Teledyne Advanced Pollution Instrumentation Automatic Zero and Span User and Communications Interfaces Modular Construction Technical Specifications Functional Description Model 633 functional block diagramEnclosure AIR Inlet and Outlet Connectors USB Connector rear panelROLL­ON and UN­ROLL Tape Spools FlowmetersDisplay and TOUCH­SCREEN USB Connectors front panelON/OFF Switch This page intentionally left blank Safety Notes and Labels Fire and explosion Instrument service and repair Moving parts Sampling line Unpacking the systemConnecting the sampling line Powering on the Aethalometer Initialization screen with test results Filter tape installation 07645A DCN6170 Teledyne Advanced Pollution Instrumentation Tape direction is from bottom, left to right Teledyne Advanced Pollution Instrumentation 07645A DCN6170 Changing filter tape Manually Lifting the Chamber Optical Head Manually lifting/lowering chamber optical headUser interface and settings User INTERFACE, Settings and Operation07645A DCN6170 Status Parameter Condition Description Instrument StatusDownloading and Viewing Data Data line HeaderData field description Connecting to External Datalogger or PC Serial Commands for Communication with the Aethalometer Supported devices External DevicesThis page intentionally left blank Frequency Maintenance and ServiceCleaning the cyclone Teledyne Advanced Pollution Instrumentation Leak test Report Automatic flow calibration procedure Automatic flow calibration procedure Automatic flow calibration procedure Manual flow meter calibration Manual flow calibration procedureManual flow calibration procedure Manual flow calibration procedure Manual flow calibration procedure Manual flow calibration procedure Manual flow calibration procedure Manual stability test Manual clean air testCleaning the optical head Optical head cleaning procedure Optical head top Optical head platform Optical head bottom Teledyne Advanced Pollution Instrumentation 07645A DCN6170 Teledyne Advanced Pollution Instrumentation Calibrating the tape sensor Teledyne Advanced Pollution Instrumentation This page intentionally left blank Technical Support and Contact Information This page intentionally left blank Spare Parts List