Teledyne 633 user manual Downloading and Viewing Data

Page 42

Model 633 Aethalometer® Black Carbon Monitor User Manual

Teledyne Advanced Pollution Instrumentation

Single Status Condition

If the value displayed matches a value in the Status Condition column, it indicates only one parameter and its description.

Examples:

Status = 0, all OK ‐> front panel LED's GREEN

Status = 1, all OK, tape advancing ‐> blinking GREEN LED

Status = 128, machine is running, tape advance warning flag is set ‐> YELLOW

Multiple Status Conditions

If the Status displayed does not match a value in the table, it means that there are multiple parameters whose Status Conditions are added together, forming a sum that must be broken down by subtraction: find the largest value in the Status Condition column that does not exceed the Status value, and subtract it from the sum. Then Find the next largest value in the Status Condition column that does not exceed the remainder, and subtract again. Continue finding the next largest number and subtracting it until the remainder matches a value in the Status Condition column.

Examples:

Status = 289, which breaks down as follows: 289 – 256 = 33; 33 – 32 = 1; therefore, the Status Conditions are 256, 32, and 1. which mean ‐> machine is not measuring (1), LED calibration had errors in 1 or more (but not all) channels (32), and less than 5 tape advances are left (256)

Status = 145, which breaks down as follows: 145 – 128 = 17; 17 – 16 = 1; therefore, the Status Conditions are 128, 17, and 1. which mean ‐> machine is not measuring (1), LED calibration is in process (16), and less than 10 tape advances are left (128)

4.3Downloading and Viewing Data

To download data, insert the USB stick in either of the front USB ports. Do not use the rear ports as they are intended for the mouse and keyboard only and not for data transfer (surge protection).

40

07645A DCN6170

Image 42
Contents Model Aethalometer Black Carbon Monitor Page Trademarks This page intentionally left blank Table of Contents Spare Parts List Introduction Description1 ‘DualSpot’ Technology Teledyne Advanced Pollution Instrumentation 07645A DCN6170 Teledyne Advanced Pollution Instrumentation Automatic Zero and Span User and Communications Interfaces Modular Construction Technical Specifications Model 633 functional block diagram Functional DescriptionEnclosure USB Connector rear panel AIR Inlet and Outlet ConnectorsFlowmeters ROLL­ON and UN­ROLL Tape SpoolsUSB Connectors front panel Display and TOUCH­SCREENON/OFF Switch This page intentionally left blank Safety Notes and Labels Fire and explosion Instrument service and repair Moving parts Unpacking the system Sampling lineConnecting the sampling line Powering on the Aethalometer Initialization screen with test results Filter tape installation 07645A DCN6170 Teledyne Advanced Pollution Instrumentation Tape direction is from bottom, left to right Teledyne Advanced Pollution Instrumentation 07645A DCN6170 Changing filter tape Manually lifting/lowering chamber optical head Manually Lifting the Chamber Optical HeadUser INTERFACE, Settings and Operation User interface and settings07645A DCN6170 Instrument Status Status Parameter Condition DescriptionDownloading and Viewing Data Header Data lineData field description Connecting to External Datalogger or PC Serial Commands for Communication with the Aethalometer External Devices Supported devicesThis page intentionally left blank Maintenance and Service FrequencyCleaning the cyclone Teledyne Advanced Pollution Instrumentation Leak test Report Automatic flow calibration procedure Automatic flow calibration procedure Automatic flow calibration procedure Manual flow calibration procedure Manual flow meter calibrationManual flow calibration procedure Manual flow calibration procedure Manual flow calibration procedure Manual flow calibration procedure Manual flow calibration procedure Manual clean air test Manual stability testCleaning the optical head Optical head cleaning procedure Optical head top Optical head platform Optical head bottom Teledyne Advanced Pollution Instrumentation 07645A DCN6170 Teledyne Advanced Pollution Instrumentation Calibrating the tape sensor Teledyne Advanced Pollution Instrumentation This page intentionally left blank Technical Support and Contact Information This page intentionally left blank Spare Parts List