5.4 Manual flow meter calibration
First, ensure that a leak test is completed (Section 5.1) prior to running this calibration procedure. For the manual calibration a flow meter, data cable and the flow calibration pad are needed. Please note flow measurements are in Milliliters, and maximum flow rate should be no greater than 5 LPM (5,000 milliliters).
Manual flow calibration procedure:
1.STOP measurements in the
OPERATION>GENERAL
menu.
2.In the
OPERATION>ADVANCED menu press FLOWCAL.
3.Select MANUAL to start the manual flow calibration.