Manual flow calibration procedure:

10.In the display’s blank FLOW field enter the flow value from your flow‐meter measured in or converted to milliliters (this will be approximately 3000; do NOT change the pressure or temperature settings), and select OK.

11.Wait for the pump to run for several seconds to achieve a stable flow.

12.In the display’s blank FLOW field enter the flow value from your flow‐meter measured in or converted to milliliters (this will be approximately 5000; do NOT change the pressure or temperature settings), and select OK.

07645A DCN6170

59

Image 61
Contents Model Aethalometer Black Carbon Monitor Page Trademarks This page intentionally left blank Table of Contents Spare Parts List Description Introduction1 ‘DualSpot’ Technology Teledyne Advanced Pollution Instrumentation 07645A DCN6170 Teledyne Advanced Pollution Instrumentation Automatic Zero and Span User and Communications Interfaces Modular Construction Technical Specifications Functional Description Model 633 functional block diagramEnclosure AIR Inlet and Outlet Connectors USB Connector rear panelROLL­ON and UN­ROLL Tape Spools FlowmetersDisplay and TOUCH­SCREEN USB Connectors front panelON/OFF Switch This page intentionally left blank Safety Notes and Labels Fire and explosion Instrument service and repair Moving parts Sampling line Unpacking the systemConnecting the sampling line Powering on the Aethalometer Initialization screen with test results Filter tape installation 07645A DCN6170 Teledyne Advanced Pollution Instrumentation Tape direction is from bottom, left to right Teledyne Advanced Pollution Instrumentation 07645A DCN6170 Changing filter tape Manually Lifting the Chamber Optical Head Manually lifting/lowering chamber optical headUser interface and settings User INTERFACE, Settings and Operation07645A DCN6170 Status Parameter Condition Description Instrument StatusDownloading and Viewing Data Data line HeaderData field description Connecting to External Datalogger or PC Serial Commands for Communication with the Aethalometer Supported devices External DevicesThis page intentionally left blank Frequency Maintenance and ServiceCleaning the cyclone Teledyne Advanced Pollution Instrumentation Leak test Report Automatic flow calibration procedure Automatic flow calibration procedure Automatic flow calibration procedure Manual flow meter calibration Manual flow calibration procedureManual flow calibration procedure Manual flow calibration procedure Manual flow calibration procedure Manual flow calibration procedure Manual flow calibration procedure Manual stability test Manual clean air testCleaning the optical head Optical head cleaning procedure Optical head top Optical head platform Optical head bottom Teledyne Advanced Pollution Instrumentation 07645A DCN6170 Teledyne Advanced Pollution Instrumentation Calibrating the tape sensor Teledyne Advanced Pollution Instrumentation This page intentionally left blank Technical Support and Contact Information This page intentionally left blank Spare Parts List