Manual flow calibration procedure:
10.In the display’s blank FLOW field enter the flow value from your flow‐meter measured in or converted to milliliters (this will be approximately 3000; do NOT change the pressure or temperature settings), and select OK.
11.Wait for the pump to run for several seconds to achieve a stable flow.
12.In the display’s blank FLOW field enter the flow value from your flow‐meter measured in or converted to milliliters (this will be approximately 5000; do NOT change the pressure or temperature settings), and select OK.