Teledyne 633 user manual Serial Commands for Communication with the Aethalometer

Page 46

Model 633 Aethalometer® Black Carbon Monitor User Manual

Teledyne Advanced Pollution Instrumentation

Establish communications using the following values:

.BaudRate = 115200

.DataBits = 8

.StopBits = IO.Ports.StopBits.One

.Parity = IO.Ports.Parity.None

.Handshake = IO.Ports.Handshake.None

.DtrEnable = True

.RtsEnable = True

4.5 Serial Commands for Communication with the Aethalometer

IMPORTANT NOTE: All commands finish with a Carriage Return [CR].

$AE33:Dnnn

nnn number of lines

Request nnn last

 

of data requested

measurements

 

 

 

 

$AE33:TyyyyMMddHHmmss

yyyy

year

Synchronize time with

 

MM

month

the data‐logger

 

dd

day

 

 

HH

hour (24)

 

 

mm

min

 

 

ss

second

 

 

 

 

 

Command: $AE33:D1[CR] will return one data line (i.e. 1 byte ‐> value = 13 so all together there should be 9 bytes)

Please note these are all capital letters.

Example 1: $AE33:D120[CR] => 11 bytes; Output will be 120 lines of data.

44

07645A DCN6170

Image 46
Contents Model Aethalometer Black Carbon Monitor Page Trademarks This page intentionally left blank Table of Contents Spare Parts List Introduction Description1 ‘DualSpot’ Technology Teledyne Advanced Pollution Instrumentation 07645A DCN6170 Teledyne Advanced Pollution Instrumentation Automatic Zero and Span User and Communications Interfaces Modular Construction Technical Specifications Model 633 functional block diagram Functional DescriptionEnclosure USB Connector rear panel AIR Inlet and Outlet ConnectorsFlowmeters ROLL­ON and UN­ROLL Tape SpoolsUSB Connectors front panel Display and TOUCH­SCREENON/OFF Switch This page intentionally left blank Safety Notes and Labels Fire and explosion Instrument service and repair Moving parts Unpacking the system Sampling lineConnecting the sampling line Powering on the Aethalometer Initialization screen with test results Filter tape installation 07645A DCN6170 Teledyne Advanced Pollution Instrumentation Tape direction is from bottom, left to right Teledyne Advanced Pollution Instrumentation 07645A DCN6170 Changing filter tape Manually lifting/lowering chamber optical head Manually Lifting the Chamber Optical HeadUser INTERFACE, Settings and Operation User interface and settings07645A DCN6170 Instrument Status Status Parameter Condition DescriptionDownloading and Viewing Data Header Data lineData field description Connecting to External Datalogger or PC Serial Commands for Communication with the Aethalometer External Devices Supported devicesThis page intentionally left blank Maintenance and Service FrequencyCleaning the cyclone Teledyne Advanced Pollution Instrumentation Leak test Report Automatic flow calibration procedure Automatic flow calibration procedure Automatic flow calibration procedure Manual flow calibration procedure Manual flow meter calibrationManual flow calibration procedure Manual flow calibration procedure Manual flow calibration procedure Manual flow calibration procedure Manual flow calibration procedure Manual clean air test Manual stability testCleaning the optical head Optical head cleaning procedure Optical head top Optical head platform Optical head bottom Teledyne Advanced Pollution Instrumentation 07645A DCN6170 Teledyne Advanced Pollution Instrumentation Calibrating the tape sensor Teledyne Advanced Pollution Instrumentation This page intentionally left blank Technical Support and Contact Information This page intentionally left blank Spare Parts List