Teledyne Advanced Pollution Instrumentation

Model 633 Aethalometer® Black Carbon Monitor User Manual

3.Disconnect flowmeter from the input.

4.Connect the flow meter to the output of the instrument by a long tube (10 m) to reduce the oscillations, which can influence the flow measurements.

5.Take a Fout reading.

IMPORTANT!

If you use a volumetric flow meter take into account the difference in air pressure and temperature between the flow in and flow out (use the ideal gas law equation).

Report

 

 

 

 

 

 

 

 

 

 

 

Aethalometer

Flowmeter

Date

Fin (LPM)

Fout (LPM)

ζ (%)

SN

 

 

 

 

 

 

 

 

 

 

 

AE33‐S00‐0054

TSI4100

15.1.2013

4.67

5.01

6.8

 

 

 

 

 

 

5.3 Automatic flow meter calibration

First, ensure that a leak test is completed (Section 5.1) prior to running this calibration procedure. For the automatic calibration a TSI4100 flow meter, data cable and the flow calibration pad are needed.

Before calibration check TSI4100 flow meter settings:

07645A DCN6170

51

Image 53
Contents Model Aethalometer Black Carbon Monitor Page Trademarks This page intentionally left blank Table of Contents Spare Parts List Description Introduction1 ‘DualSpot’ Technology Teledyne Advanced Pollution Instrumentation 07645A DCN6170 Teledyne Advanced Pollution Instrumentation Automatic Zero and Span User and Communications Interfaces Modular Construction Technical Specifications Functional Description Model 633 functional block diagramEnclosure AIR Inlet and Outlet Connectors USB Connector rear panelROLL­ON and UN­ROLL Tape Spools FlowmetersDisplay and TOUCH­SCREEN USB Connectors front panelON/OFF Switch This page intentionally left blank Safety Notes and Labels Fire and explosion Instrument service and repair Moving parts Sampling line Unpacking the systemConnecting the sampling line Powering on the Aethalometer Initialization screen with test results Filter tape installation 07645A DCN6170 Teledyne Advanced Pollution Instrumentation Tape direction is from bottom, left to right Teledyne Advanced Pollution Instrumentation 07645A DCN6170 Changing filter tape Manually Lifting the Chamber Optical Head Manually lifting/lowering chamber optical headUser interface and settings User INTERFACE, Settings and Operation07645A DCN6170 Status Parameter Condition Description Instrument StatusDownloading and Viewing Data Data line HeaderData field description Connecting to External Datalogger or PC Serial Commands for Communication with the Aethalometer Supported devices External DevicesThis page intentionally left blank Frequency Maintenance and ServiceCleaning the cyclone Teledyne Advanced Pollution Instrumentation Leak test Report Automatic flow calibration procedure Automatic flow calibration procedure Automatic flow calibration procedure Manual flow meter calibration Manual flow calibration procedureManual flow calibration procedure Manual flow calibration procedure Manual flow calibration procedure Manual flow calibration procedure Manual flow calibration procedure Manual stability test Manual clean air testCleaning the optical head Optical head cleaning procedure Optical head top Optical head platform Optical head bottom Teledyne Advanced Pollution Instrumentation 07645A DCN6170 Teledyne Advanced Pollution Instrumentation Calibrating the tape sensor Teledyne Advanced Pollution Instrumentation This page intentionally left blank Technical Support and Contact Information This page intentionally left blank Spare Parts List