Model 633 Aethalometer® Black Carbon Monitor User Manual | Teledyne Advanced Pollution Instrumentation |
Manual flow calibration procedure:
8.In the display’s blank FLOW field enter the flow value from your flow meter measured in or converted to milliliters (this will be approximately 900) tap screen in Flow field, used screen keyboard to enter numbers and click Enter.
Also enter the EPA standard temp and pressure of 25 C and 1 atmosphere or 101325 Pa and select OK.
9.Wait for the pump to run for several seconds to achieve a stable flow.