Installation - 2

Input Connections

Input Source and Line Fuse

The ac source requires a 3-phase power service that provides 7350 VA (6000 W) maximum. The power service should have a current rating greater than or equal to the ac source’s circuit breaker rating. The ac source has a delta input (no neutral connection) and will accept power from either delta (triangle) or wye (star) services.

CAUTION: Two input voltage ranges are available (see "AC Input Ratings" in appendix A). The ac source will be damaged if it is operated at an input voltage that is outside of its configured input range.

In order to maintain phase current balancing, the power service should be a dedicated line with only Agilent 6814B/6834B/6843A ac sources drawing current from it. A disconnect box located near the ac source is recommended for all installations and is mandatory for direct-wired installations.

Installing the Power Cord

WARNING: Installation of the power cord must be done by a qualified and licensed electrician and must be in accordance with local electrical codes.

The power cords supplied with the ac source do not include a power plug. Terminating connectors and a ground lug are attached to one end of the cord. See Figure 2-2 while performing the following procedure.

1.Check the line fuses as follows:

a.Examine the FUSES label on the rear panel.

b.Remove the safety cover in front of the fuse caps.

c.Unscrew the line fuse caps from the rear panel and verify that all fuses are as specified on the label. Reinstall the fuses.

d.Reinstall the safety cover in front of the fuse caps.

2.Remove the ac input cover from the back of the unit.

3.Open the line clamp on the ac input cover and insert the line cord through the opening.

4.Position the power cord so that the clamp is near the end of the outside insulating sheath on the power cord. Tighten the screws to secure the clamp.

5.Secure the three ac lines to the ac power strip as follows:

Phase 1 to L1 (black).

Phase 2 to L2 (red).

Phase 3 to L3 (orange).

6.Secure the ground wire (green) to the chassis earth ground terminal.

7.

Slip the safety cover over the ac input terminal strip and secure the cover with the four

 

screws.

8.If required, wire the appropriate power plug to the other end of the power cord. For user- made cables, strip back the sheath 10 cm (4 in).

17

Image 17
Contents User’s Guide AC Power Solutions Agilent Models 6814B, 6834B, and 6843ACertification Warranty InformationGeneral Safety SummarySafety Symbols EMC DeclarationPrinting History Acoustic Noise InformationTable of Contents Performing the Calibration Procedure Specifications Supplemental CharacteristicsError Number List Entry Keys Examples of Front Panel ProgrammingLocation Document OrientationTopic Options, Accessories, and User Replaceable Parts Safety ConsiderationsOption Description Agilent Part NumberModel Description DescriptionCapabilities Ranges Output CharacteristicFront Panel/Remote Operation Output VA Capability AC Source Output Characteristic in real-time modePage Inspection CleaningDamage Packaging Material Items SuppliedBench Operation LocationRack Mounting Not block the fan exhaust at the rear of the unitInput Source and Line Fuse Installing the Power CordInput Connections Output Connections Output ConnectionsVoltage Drops Wire ConsiderationsCurrent Ratings Ampacity and Resistance of Stranded Copper ConductorsRemote Sense Connections Remote Sense ConnectionsDigital Connections Trigger ConnectionsOVP Considerations Output RatingController Connections Gpib ConnectorRS-232 Connector Interface CommandsRS-232 Interface RS-232 Data FormatNull Modem Interface Lines Response Data Terminator Hardware HandshakePreliminary Checkout IntroductionCheckout Procedure Using the KeypadProtclear Procedure Display ExplanationVolt Currlev VoltprotCurrprot Currprot OFFLine Fuse Error MessagesCase of Trouble Page Front Panel, Overall View Front Panel DescriptionAC+DC System Keys Display Command FunctionImmediate Action Keys Annunciator On Phase SelectedFunction Keys Meter Display Keys Scrolling KeysDisplay Measurement KBESSEL, RectDisplay Output Control KeysRST CLS Protection and Status Control KeysInitimmed Trigger and List Control KeysIMM AbortThrough Are used for entering numeric values Is the decimal Entry KeysProcedure for Single-Phase AC Sources Setting the Output Voltage AmplitudeExamples of Front Panel Programming Set the output to 120 V rms as followsPhase 3 to 235 Vrms Procedure for Three-Phase AC SourcesAction Display To verify the output, you can measure it as followsOvercurrent protection feature as follows Setting the Output FrequencySetting a Protection Feature Step Transient Using Transient Voltage ModesVoltt Voltm StepVoltm Pulse Pulse TransientWidth DcycleVoltm Fixed Voltm List Count List TransientAction Volt EOL Step Auto Trigger Delays and Phase SynchronizationSyncphas Initimmed Syncsour PhaseExample Display Example Voltm Step Syncsour Phase Syncphas InitimmedProgramming Slew Rates Using Slew Rates to Generate WaveformsSlew Voltm FixedStep Slewm SlewtMeasuring Peak Inrush Current Syncsour Phase SyncphasRange CurrpeakTo configure the RS-232 interface, proceed as follows Setting the Gpib Address and RS-232 ParametersSaving and Recalling Operating States Action Display To set the Gpib address, proceed as followsAddress Noutputs Agilent 6814B Agilent 6834B Agilent 6843A SpecificationsTable A-1. Performance Specifications1 Table A-2. Supplemental Characteristics Supplemental CharacteristicsSpecifications a Page Recommended Model CharacteristicsEquipment Required Performing the Verification Tests Test SetupTurn-On Checkout Procedure Action Normal Result Voltage Programming and Measurement AccuracyRMS Current Readback Accuracy Agilent 6814B Current Measurement Accuracy Performing the Calibration ProcedureAgilent 6834B Current Measurement Accuracy Agilent 6843A Current Measurement AccuracyFront Panel Calibration Menu Enable Calibration ModeFront Panel Calibration Calibrating the OVP trip point Calibrating and Entering Voltage Calibration ValuesCalibrating and Entering Current Calibration Values Calibrating the Output Impedance Agilent 6843A only Changing the Calibration PasswordSaving the Calibration Constants CalpassTable B-3. Gpib Calibration Error Messages Calibration Error MessagesCalibration Over the Gpib Agilent Calibration Program ListingFigure B-2. Calibration Program Listing Sheet 1 Figure B-2. Calibration Program Listing Sheet 2 Page Table C-1. Error Numbers Error Number ListError Messages Error Messages C Page Index Index Index Canada Australia/New Zealand United States Latin AmericaEurope Asia Pacific JapanManual Updates
Related manuals
Manual 20 pages 27.01 Kb Manual 187 pages 38.86 Kb

6834B, 6843A, 6814B specifications

Agilent Technologies, a leader in electronic test and measurement equipment, offers a range of powerful signal sources including the 6843A, 6834B, 6814B, 6813B, and 6811B models. These instruments are designed to support various applications in research, development, and manufacturing, providing precise signal generation capabilities.

The Agilent 6843A is a versatile signal generator known for its exceptional frequency range and modulation capabilities. It supports an extensive bandwidth, making it ideal for applications that require high-frequency signal generation. With its superior phase noise performance, the 6843A is an excellent choice for radar, wireless communications, and electronic warfare applications. The instrument features an intuitive user interface, allowing engineers to set parameters quickly and efficiently.

Next, the Agilent 6834B offers exceptional performance characteristics, including high output power and low distortion. This signal generator is particularly noted for its ability to produce complex modulation formats, making it suitable for testing advanced wireless communication systems. With a reliable and stable output, the 6834B ensures accurate and repeatable measurements, which is vital for thorough testing processes.

The 6814B model stands out for its dual-channel capabilities, allowing users to generate simultaneous signals for testing multiple components or systems. This feature significantly enhances testing efficiency and flexibility for engineers. With built-in arbitrary waveform functionality, users can create custom waveforms, making the 6814B suitable for a wide range of applications including device characterization and signal processing research.

For those seeking a more compact solution, the Agilent 6813B provides essential signal generation features without compromising on performance. It is designed for a variety of applications across telecommunications and consumer electronics, featuring a straightforward interface and robust performance metrics.

Lastly, the 6811B is an entry-level yet capable model that supports a broad spectrum of testing needs. Perfect for educational and laboratory environments, it provides essential functionalities required for effective signal generation and analysis.

Overall, Agilent Technologies' 6843A, 6834B, 6814B, 6813B, and 6811B signal generators offer an array of features and technologies that cater to various application needs. Their precision, reliability, and user-oriented designs position them as invaluable assets in any testing environment, ensuring engineers can carry out their work with confidence and accuracy.