Agilent Technologies 6834B, 6814B, 6843A manual Safety Summary, General

Page 3

Safety Summary

The following general safety precautions must be observed during all phases of operation of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements.

WARNING: LETHAL VOLTAGES

Ac sources can supply 425 V peak at their output. DEATH on contact may result if the output terminals or circuits connected to the output are touched when power is applied.

GENERAL

This product is a Safety Class 1 instrument (provided with a protective earth terminal). The protective features of this product may be impaired if it is used in a manner not specified in the operation instructions.

Any LEDs used in this product are Class 1 LEDs as per IEC 825-1.

ENVIRONMENTAL CONDITONS

This instrument is intended for indoor use in an installation category III, pollution degree 2 environment. It is designed to operate at a maximum relative humidity of 95% and at altitudes of up to 2000 meters. Refer to the specifications tables for the ac mains voltage requirements and ambient operating temperature range.

BEFORE APPLYING POWER

Verify that the product is set to match the available line voltage, the correct fuse is installed, and all safety precautions are taken. Note the instrument’s external markings described under "Safety Symbols".

GROUND THE INSTRUMENT

To minimize shock hazard, the instrument chassis and cover must be connected to an electrical ground. The instrument must be connected to the ac power mains through a grounded power cable, with the ground wire firmly connected to an electrical ground (safety ground) at the power outlet. Any interruption of the protective (grounding) conductor or disconnection of the protective earth terminal will cause a potential shock hazard that could result in personal injury.

ATTENTION: Un circuit de terre continu est essentiel en vue du fonctionnement sécuritaire de l’appareil. Ne jamais mettre l'appareil en marche lorsque le conducteur de mise … la terre est d‚branch‚.

FUSES

Only fuses with the required rated current, voltage, and specified type (normal blow, time delay, etc.) should be used. Do not use repaired fuses or short-circuited fuseholders. To do so could cause a shock or fire hazard.

DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE

Do not operate the instrument in the presence of flammable gases or fumes.

DO NOT REMOVE THE INSTRUMENT COVER

Operating personnel must not remove instrument covers. Component replacement and internal adjustments must be made only by qualified service personnel.

DO NOT EXCEED INPUT RATINGS

Operation at line voltages or frequencies in excess of those stated on the line rating label may cause leakage currents in excess of 5.0 mA peak.

Instruments that appear damaged or defective should be made inoperative and secured against unintended operation until they can be repaired by qualified service personnel.

3

Image 3
Contents User’s Guide AC Power Solutions Agilent Models 6814B, 6834B, and 6843ACertification Warranty InformationGeneral Safety SummarySafety Symbols EMC DeclarationPrinting History Acoustic Noise InformationTable of Contents Entry Keys Examples of Front Panel Programming Specifications Supplemental CharacteristicsPerforming the Calibration Procedure Error Number ListDocument Orientation TopicLocation Agilent Part Number Safety ConsiderationsOptions, Accessories, and User Replaceable Parts Option DescriptionDescription CapabilitiesModel Description Output Characteristic Front Panel/Remote OperationRanges Output VA Capability AC Source Output Characteristic in real-time modePage Items Supplied CleaningInspection Damage Packaging MaterialNot block the fan exhaust at the rear of the unit LocationBench Operation Rack MountingInstalling the Power Cord Input ConnectionsInput Source and Line Fuse Output Connections Output ConnectionsAmpacity and Resistance of Stranded Copper Conductors Wire ConsiderationsVoltage Drops Current RatingsRemote Sense Connections Remote Sense ConnectionsOutput Rating Trigger ConnectionsDigital Connections OVP ConsiderationsController Connections Gpib ConnectorRS-232 Data Format Interface CommandsRS-232 Connector RS-232 InterfaceNull Modem Interface Lines Response Data Terminator Hardware HandshakePreliminary Checkout IntroductionCheckout Procedure Using the KeypadProcedure Display Explanation VoltProtclear Currprot OFF VoltprotCurrlev CurrprotError Messages Case of TroubleLine Fuse Page Front Panel, Overall View Front Panel DescriptionAC+DC System Keys Display Command FunctionAnnunciator On Phase Selected Function KeysImmediate Action Keys KBESSEL, Rect Scrolling KeysMeter Display Keys Display MeasurementDisplay Output Control KeysRST CLS Protection and Status Control KeysAbort Trigger and List Control KeysInitimmed IMMThrough Are used for entering numeric values Is the decimal Entry KeysSet the output to 120 V rms as follows Setting the Output Voltage AmplitudeProcedure for Single-Phase AC Sources Examples of Front Panel ProgrammingTo verify the output, you can measure it as follows Procedure for Three-Phase AC SourcesPhase 3 to 235 Vrms Action DisplaySetting the Output Frequency Setting a Protection FeatureOvercurrent protection feature as follows Voltm Step Using Transient Voltage ModesStep Transient VolttDcycle Pulse TransientVoltm Pulse WidthList Transient ActionVoltm Fixed Voltm List Count Volt EOL Step Auto Trigger Delays and Phase SynchronizationSyncsour Phase Example DisplaySyncphas Initimmed Example Voltm Step Syncsour Phase Syncphas InitimmedProgramming Slew Rates Using Slew Rates to Generate WaveformsSlewt Voltm FixedSlew Step SlewmCurrpeak Syncsour Phase SyncphasMeasuring Peak Inrush Current RangeAction Display To set the Gpib address, proceed as follows Setting the Gpib Address and RS-232 ParametersTo configure the RS-232 interface, proceed as follows Saving and Recalling Operating StatesAddress Noutputs Specifications Table A-1. Performance Specifications1Agilent 6814B Agilent 6834B Agilent 6843A Table A-2. Supplemental Characteristics Supplemental CharacteristicsSpecifications a Page Characteristics Equipment RequiredRecommended Model Test Setup Turn-On Checkout ProcedurePerforming the Verification Tests Voltage Programming and Measurement Accuracy RMS Current Readback AccuracyAction Normal Result Agilent 6843A Current Measurement Accuracy Performing the Calibration ProcedureAgilent 6814B Current Measurement Accuracy Agilent 6834B Current Measurement AccuracyEnable Calibration Mode Front Panel CalibrationFront Panel Calibration Menu Calibrating the OVP trip point Calibrating and Entering Voltage Calibration ValuesCalibrating and Entering Current Calibration Values Calpass Changing the Calibration PasswordCalibrating the Output Impedance Agilent 6843A only Saving the Calibration ConstantsAgilent Calibration Program Listing Calibration Error MessagesTable B-3. Gpib Calibration Error Messages Calibration Over the GpibFigure B-2. Calibration Program Listing Sheet 1 Figure B-2. Calibration Program Listing Sheet 2 Page Table C-1. Error Numbers Error Number ListError Messages Error Messages C Page Index Index Index Japan United States Latin AmericaCanada Australia/New Zealand Europe Asia PacificManual Updates
Related manuals
Manual 20 pages 27.01 Kb Manual 187 pages 38.86 Kb

6834B, 6843A, 6814B specifications

Agilent Technologies, a leader in electronic test and measurement equipment, offers a range of powerful signal sources including the 6843A, 6834B, 6814B, 6813B, and 6811B models. These instruments are designed to support various applications in research, development, and manufacturing, providing precise signal generation capabilities.

The Agilent 6843A is a versatile signal generator known for its exceptional frequency range and modulation capabilities. It supports an extensive bandwidth, making it ideal for applications that require high-frequency signal generation. With its superior phase noise performance, the 6843A is an excellent choice for radar, wireless communications, and electronic warfare applications. The instrument features an intuitive user interface, allowing engineers to set parameters quickly and efficiently.

Next, the Agilent 6834B offers exceptional performance characteristics, including high output power and low distortion. This signal generator is particularly noted for its ability to produce complex modulation formats, making it suitable for testing advanced wireless communication systems. With a reliable and stable output, the 6834B ensures accurate and repeatable measurements, which is vital for thorough testing processes.

The 6814B model stands out for its dual-channel capabilities, allowing users to generate simultaneous signals for testing multiple components or systems. This feature significantly enhances testing efficiency and flexibility for engineers. With built-in arbitrary waveform functionality, users can create custom waveforms, making the 6814B suitable for a wide range of applications including device characterization and signal processing research.

For those seeking a more compact solution, the Agilent 6813B provides essential signal generation features without compromising on performance. It is designed for a variety of applications across telecommunications and consumer electronics, featuring a straightforward interface and robust performance metrics.

Lastly, the 6811B is an entry-level yet capable model that supports a broad spectrum of testing needs. Perfect for educational and laboratory environments, it provides essential functionalities required for effective signal generation and analysis.

Overall, Agilent Technologies' 6843A, 6834B, 6814B, 6813B, and 6811B signal generators offer an array of features and technologies that cater to various application needs. Their precision, reliability, and user-oriented designs position them as invaluable assets in any testing environment, ensuring engineers can carry out their work with confidence and accuracy.