Agilent Technologies 6834B, 6814B, 6843A Front Panel Calibration Menu, Enable Calibration Mode

Page 63

Verification and Calibration - B

Front Panel Calibration Menu

The Entry keypad is used for calibration functions.

Shift

Cal

Press this key to access the calibration menu.

Display

Command Function

CAL ON <value>

Turns calibration mode on when the correct password

 

value is entered.

CAL OFF

Turns calibration mode off

CAL:LEV <char>

Advance to next step in sequence (P1, P2, P3, or P4).

CAL:DATA <value>

Input a calibration measurement.

CAL:VOLT:AC

Begin ac voltage calibration sequence

CAL:VOLT:PROT

Begin voltage protection calibration

CAL:CURR:AC

Begin ac current calibration sequence

CAL:CURR:MEAS

Begin current measurement calibration sequence

CAL:IMP

Begin output impedance calibration sequence

CAL:VOLT:RTIM1

Begin realtime voltage calibration

CAL:SAVE

Saves the calibration constants in non-volatile memory.

CAL:PASS <value>

Set new calibration password.

Notes:

1 Agilent 6843A only

 

value = a numeric value

char = a character string parameter

Use p

and

q

to scroll through the command list.

Use À

and

Ã

to scroll through the parameter list.

Front Panel Calibration

WARNING LETHAL VOLTAGES. Ac sources can supply 425 V peak at their output. DEATH on contact may result if the output terminals or circuits connected to the output are touched when power is applied. These procedures must be performed by a qualified electronics technician or engineer trained on this equipment.

The following procedures assume you understand how to operate front panel keys (see Chapter 4).

Enable Calibration Mode

 

Action

Display

1.

Reset the unit by selecting Shift, Output, and pressing Enter.

*RST

 

If you are calibration a 3-phase source , make sure that the default phase 1

 

 

(φ1) annunciator is lit. Press Phase Select to select a different phase.

 

2.

Press Shift Calibration, scroll to CAL ON, and press Enter.

CAL ON 0.0

3.Enter the calibration password from Entry keypad and press Enter. If the password is correct the Cal annunciator will come on.

If CAL DENIED appears, then an internal switch has been set to prevent the calibration from being changed. (see the Service Manual.)

If the password is incorrect, an error occurs. If the active password is lost, the calibration function can be recovered by setting an internal switch that defeats password protection. (see the Service Manual.)

CAL DENIED

OUT OF RANGE

63

Image 63
Contents User’s Guide AC Power Solutions Agilent Models 6814B, 6834B, and 6843ACertification Warranty InformationGeneral Safety SummarySafety Symbols EMC DeclarationPrinting History Acoustic Noise InformationTable of Contents Entry Keys Examples of Front Panel Programming Specifications Supplemental CharacteristicsPerforming the Calibration Procedure Error Number ListDocument Orientation TopicLocation Agilent Part Number Safety ConsiderationsOptions, Accessories, and User Replaceable Parts Option DescriptionDescription CapabilitiesModel Description Output Characteristic Front Panel/Remote OperationRanges Output VA Capability AC Source Output Characteristic in real-time modePage Items Supplied CleaningInspection Damage Packaging MaterialNot block the fan exhaust at the rear of the unit LocationBench Operation Rack MountingInstalling the Power Cord Input ConnectionsInput Source and Line Fuse Output Connections Output ConnectionsAmpacity and Resistance of Stranded Copper Conductors Wire ConsiderationsVoltage Drops Current RatingsRemote Sense Connections Remote Sense ConnectionsOutput Rating Trigger ConnectionsDigital Connections OVP ConsiderationsController Connections Gpib ConnectorRS-232 Data Format Interface CommandsRS-232 Connector RS-232 InterfaceNull Modem Interface Lines Response Data Terminator Hardware HandshakePreliminary Checkout IntroductionCheckout Procedure Using the KeypadProcedure Display Explanation VoltProtclear Currprot OFF VoltprotCurrlev CurrprotError Messages Case of TroubleLine Fuse Page Front Panel, Overall View Front Panel DescriptionAC+DC System Keys Display Command FunctionAnnunciator On Phase Selected Function KeysImmediate Action Keys KBESSEL, Rect Scrolling KeysMeter Display Keys Display MeasurementDisplay Output Control KeysRST CLS Protection and Status Control KeysAbort Trigger and List Control KeysInitimmed IMMThrough Are used for entering numeric values Is the decimal Entry KeysSet the output to 120 V rms as follows Setting the Output Voltage AmplitudeProcedure for Single-Phase AC Sources Examples of Front Panel ProgrammingTo verify the output, you can measure it as follows Procedure for Three-Phase AC SourcesPhase 3 to 235 Vrms Action DisplaySetting the Output Frequency Setting a Protection FeatureOvercurrent protection feature as follows Voltm Step Using Transient Voltage ModesStep Transient VolttDcycle Pulse TransientVoltm Pulse WidthList Transient ActionVoltm Fixed Voltm List Count Volt EOL Step Auto Trigger Delays and Phase SynchronizationSyncsour Phase Example DisplaySyncphas Initimmed Example Voltm Step Syncsour Phase Syncphas InitimmedProgramming Slew Rates Using Slew Rates to Generate WaveformsSlewt Voltm FixedSlew Step SlewmCurrpeak Syncsour Phase SyncphasMeasuring Peak Inrush Current RangeAction Display To set the Gpib address, proceed as follows Setting the Gpib Address and RS-232 ParametersTo configure the RS-232 interface, proceed as follows Saving and Recalling Operating StatesAddress Noutputs Specifications Table A-1. Performance Specifications1Agilent 6814B Agilent 6834B Agilent 6843A Table A-2. Supplemental Characteristics Supplemental CharacteristicsSpecifications a Page Characteristics Equipment RequiredRecommended Model Test Setup Turn-On Checkout ProcedurePerforming the Verification Tests Voltage Programming and Measurement Accuracy RMS Current Readback AccuracyAction Normal Result Agilent 6843A Current Measurement Accuracy Performing the Calibration ProcedureAgilent 6814B Current Measurement Accuracy Agilent 6834B Current Measurement AccuracyEnable Calibration Mode Front Panel CalibrationFront Panel Calibration Menu Calibrating the OVP trip point Calibrating and Entering Voltage Calibration ValuesCalibrating and Entering Current Calibration Values Calpass Changing the Calibration PasswordCalibrating the Output Impedance Agilent 6843A only Saving the Calibration ConstantsAgilent Calibration Program Listing Calibration Error MessagesTable B-3. Gpib Calibration Error Messages Calibration Over the GpibFigure B-2. Calibration Program Listing Sheet 1 Figure B-2. Calibration Program Listing Sheet 2 Page Table C-1. Error Numbers Error Number ListError Messages Error Messages C Page Index Index Index Japan United States Latin AmericaCanada Australia/New Zealand Europe Asia PacificManual Updates
Related manuals
Manual 20 pages 27.01 Kb Manual 187 pages 38.86 Kb

6834B, 6843A, 6814B specifications

Agilent Technologies, a leader in electronic test and measurement equipment, offers a range of powerful signal sources including the 6843A, 6834B, 6814B, 6813B, and 6811B models. These instruments are designed to support various applications in research, development, and manufacturing, providing precise signal generation capabilities.

The Agilent 6843A is a versatile signal generator known for its exceptional frequency range and modulation capabilities. It supports an extensive bandwidth, making it ideal for applications that require high-frequency signal generation. With its superior phase noise performance, the 6843A is an excellent choice for radar, wireless communications, and electronic warfare applications. The instrument features an intuitive user interface, allowing engineers to set parameters quickly and efficiently.

Next, the Agilent 6834B offers exceptional performance characteristics, including high output power and low distortion. This signal generator is particularly noted for its ability to produce complex modulation formats, making it suitable for testing advanced wireless communication systems. With a reliable and stable output, the 6834B ensures accurate and repeatable measurements, which is vital for thorough testing processes.

The 6814B model stands out for its dual-channel capabilities, allowing users to generate simultaneous signals for testing multiple components or systems. This feature significantly enhances testing efficiency and flexibility for engineers. With built-in arbitrary waveform functionality, users can create custom waveforms, making the 6814B suitable for a wide range of applications including device characterization and signal processing research.

For those seeking a more compact solution, the Agilent 6813B provides essential signal generation features without compromising on performance. It is designed for a variety of applications across telecommunications and consumer electronics, featuring a straightforward interface and robust performance metrics.

Lastly, the 6811B is an entry-level yet capable model that supports a broad spectrum of testing needs. Perfect for educational and laboratory environments, it provides essential functionalities required for effective signal generation and analysis.

Overall, Agilent Technologies' 6843A, 6834B, 6814B, 6813B, and 6811B signal generators offer an array of features and technologies that cater to various application needs. Their precision, reliability, and user-oriented designs position them as invaluable assets in any testing environment, ensuring engineers can carry out their work with confidence and accuracy.