Agilent Technologies 6834B, 6814B Trigger Connections, Digital Connections, OVP Considerations

Page 21

Installation - 2

OVP Considerations

The overvoltage protection circuit senses voltage near the output terminals, not at the load. Therefore the signal sensed by the OVP circuit can be significantly higher than the actual voltage at the load. When using remote sensing, you must program the OVP trip voltage high enough to compensate for the voltage drop between the output terminals and the load.

Output Rating

In remote sense applications, the voltage drop in the load leads subtracts from the available load voltage (see "Remote Sensing Capability" in appendix A). As the ac source increases its output to overcome this voltage drop, the sum of the programmed voltage and the load-lead drop may exceed the ac source’s maximum voltage rating. This will not damage the unit, but may trip the OV protection circuit, which senses the voltage at the output terminals.

Trigger Connections

The BNC trigger connectors on the rear panel let you apply trigger signals to the ac source as well as generate trigger signals from the ac source. The electrical characteristics of the trigger connectors are described in appendix A. More information on programming external triggers is found in Chapter 4 of the ac source Programming Guide.

Trigger IN Allows negative-going external trigger signals to trigger the ac source.

Trigger OUT Generates a negative-going pulse when the selected transient output has occurred.

Digital Connections

This connector, which is on the rear panel, is for connecting the fault and the inhibit signals. The fault (FLT) signal is also referred to as the DFI signal in the front panel and SCPI commands. The inhibit

(INH) signal is also referred to as the RI signal in the front panel and SCPI commands.

The connector accepts wires sizes from AWG 22 to AWG 12. Disconnect the mating plug to make your wire connections. The electrical characteristics of the digital connectors are described in appendix A. More information on programming the digital connectors is found in Chapter 4 of the ac source Programming Guide.

NOTE: It is good engineering practice to twist and shield all signal wires to and from the digital connectors

The following examples show how you can connect the FLT/INH circuits of the ac source.

In example A, the INH input connects to a switch that shorts pin + to pin ⊥ whenever it is necessary to disable output of the unit. This activates the remote inhibit (RI) circuit, which turns off the ac output. The front panel Prot annunciator comes on and the RI bit is set in the Questionable Status Event register. To re-enable the unit, first open the connection between pins + and ⊥ and then clear the protection circuit. This can be done either from the front panel or over the GPIB /RS-232.

21

Image 21
Contents User’s Guide AC Power Solutions Agilent Models 6814B, 6834B, and 6843ACertification Warranty InformationGeneral Safety SummarySafety Symbols EMC DeclarationPrinting History Acoustic Noise InformationTable of Contents Performing the Calibration Procedure Specifications Supplemental CharacteristicsError Number List Entry Keys Examples of Front Panel ProgrammingDocument Orientation TopicLocation Options, Accessories, and User Replaceable Parts Safety ConsiderationsOption Description Agilent Part NumberDescription CapabilitiesModel Description Output Characteristic Front Panel/Remote OperationRanges Output VA Capability AC Source Output Characteristic in real-time modePage Inspection CleaningDamage Packaging Material Items SuppliedBench Operation LocationRack Mounting Not block the fan exhaust at the rear of the unitInstalling the Power Cord Input ConnectionsInput Source and Line Fuse Output Connections Output ConnectionsVoltage Drops Wire ConsiderationsCurrent Ratings Ampacity and Resistance of Stranded Copper ConductorsRemote Sense Connections Remote Sense ConnectionsDigital Connections Trigger ConnectionsOVP Considerations Output RatingController Connections Gpib ConnectorRS-232 Connector Interface CommandsRS-232 Interface RS-232 Data FormatNull Modem Interface Lines Response Data Terminator Hardware HandshakePreliminary Checkout IntroductionCheckout Procedure Using the KeypadProcedure Display Explanation VoltProtclear Currlev VoltprotCurrprot Currprot OFFError Messages Case of TroubleLine Fuse Page Front Panel, Overall View Front Panel DescriptionAC+DC System Keys Display Command FunctionAnnunciator On Phase Selected Function KeysImmediate Action Keys Meter Display Keys Scrolling KeysDisplay Measurement KBESSEL, RectDisplay Output Control KeysRST CLS Protection and Status Control KeysInitimmed Trigger and List Control KeysIMM AbortThrough Are used for entering numeric values Is the decimal Entry KeysProcedure for Single-Phase AC Sources Setting the Output Voltage AmplitudeExamples of Front Panel Programming Set the output to 120 V rms as followsPhase 3 to 235 Vrms Procedure for Three-Phase AC SourcesAction Display To verify the output, you can measure it as followsSetting the Output Frequency Setting a Protection FeatureOvercurrent protection feature as follows Step Transient Using Transient Voltage ModesVoltt Voltm StepVoltm Pulse Pulse TransientWidth DcycleList Transient ActionVoltm Fixed Voltm List Count Volt EOL Step Auto Trigger Delays and Phase SynchronizationSyncsour Phase Example DisplaySyncphas Initimmed Example Voltm Step Syncsour Phase Syncphas InitimmedProgramming Slew Rates Using Slew Rates to Generate WaveformsSlew Voltm FixedStep Slewm SlewtMeasuring Peak Inrush Current Syncsour Phase SyncphasRange CurrpeakTo configure the RS-232 interface, proceed as follows Setting the Gpib Address and RS-232 ParametersSaving and Recalling Operating States Action Display To set the Gpib address, proceed as followsAddress Noutputs Specifications Table A-1. Performance Specifications1Agilent 6814B Agilent 6834B Agilent 6843A Table A-2. Supplemental Characteristics Supplemental CharacteristicsSpecifications a Page Characteristics Equipment RequiredRecommended Model Test Setup Turn-On Checkout ProcedurePerforming the Verification Tests Voltage Programming and Measurement Accuracy RMS Current Readback AccuracyAction Normal Result Agilent 6814B Current Measurement Accuracy Performing the Calibration ProcedureAgilent 6834B Current Measurement Accuracy Agilent 6843A Current Measurement AccuracyEnable Calibration Mode Front Panel CalibrationFront Panel Calibration Menu Calibrating the OVP trip point Calibrating and Entering Voltage Calibration ValuesCalibrating and Entering Current Calibration Values Calibrating the Output Impedance Agilent 6843A only Changing the Calibration PasswordSaving the Calibration Constants CalpassTable B-3. Gpib Calibration Error Messages Calibration Error MessagesCalibration Over the Gpib Agilent Calibration Program ListingFigure B-2. Calibration Program Listing Sheet 1 Figure B-2. Calibration Program Listing Sheet 2 Page Table C-1. Error Numbers Error Number ListError Messages Error Messages C Page Index Index Index Canada Australia/New Zealand United States Latin AmericaEurope Asia Pacific JapanManual Updates
Related manuals
Manual 20 pages 27.01 Kb Manual 187 pages 38.86 Kb

6834B, 6843A, 6814B specifications

Agilent Technologies, a leader in electronic test and measurement equipment, offers a range of powerful signal sources including the 6843A, 6834B, 6814B, 6813B, and 6811B models. These instruments are designed to support various applications in research, development, and manufacturing, providing precise signal generation capabilities.

The Agilent 6843A is a versatile signal generator known for its exceptional frequency range and modulation capabilities. It supports an extensive bandwidth, making it ideal for applications that require high-frequency signal generation. With its superior phase noise performance, the 6843A is an excellent choice for radar, wireless communications, and electronic warfare applications. The instrument features an intuitive user interface, allowing engineers to set parameters quickly and efficiently.

Next, the Agilent 6834B offers exceptional performance characteristics, including high output power and low distortion. This signal generator is particularly noted for its ability to produce complex modulation formats, making it suitable for testing advanced wireless communication systems. With a reliable and stable output, the 6834B ensures accurate and repeatable measurements, which is vital for thorough testing processes.

The 6814B model stands out for its dual-channel capabilities, allowing users to generate simultaneous signals for testing multiple components or systems. This feature significantly enhances testing efficiency and flexibility for engineers. With built-in arbitrary waveform functionality, users can create custom waveforms, making the 6814B suitable for a wide range of applications including device characterization and signal processing research.

For those seeking a more compact solution, the Agilent 6813B provides essential signal generation features without compromising on performance. It is designed for a variety of applications across telecommunications and consumer electronics, featuring a straightforward interface and robust performance metrics.

Lastly, the 6811B is an entry-level yet capable model that supports a broad spectrum of testing needs. Perfect for educational and laboratory environments, it provides essential functionalities required for effective signal generation and analysis.

Overall, Agilent Technologies' 6843A, 6834B, 6814B, 6813B, and 6811B signal generators offer an array of features and technologies that cater to various application needs. Their precision, reliability, and user-oriented designs position them as invaluable assets in any testing environment, ensuring engineers can carry out their work with confidence and accuracy.