Agilent Technologies 6814B Voltage Programming and Measurement Accuracy, Action Normal Result

Page 61

Verification and Calibration - B

Voltage Programming and Measurement Accuracy

This test verifies the voltage programming, GPIB measurement, and front panel meter functions. Values read back over the GPIB should be the same as those displayed on the front panel.

Figure B-1 shows the setup. Measure the ac output voltage directly at the output terminals. If you are verifying a three-phase source, sart by verifying output phase 1.

Action

Normal Result

1.Make sure the ac source is turned off. Connect the DVM and ratio transformer as shown in the test setup in Figure B-1.

2.

Turn on the ac source with no load. In the Output menu, execute the

*RST

 

*RST command to reset the unit to its factory default state.

 

3.

Program the output voltage to 150 volts and set the output current

CV annunciator on.

 

limit to its maximum value.

Output voltage near 0.

 

 

Output current near 0.

4.

Enable the output by pressing Output On/Off.

Output voltage near 150 V.

5.

Record voltage readings at the DVM1 and on the front panel display.

Readings within low voltage

 

 

limits specified in table B-2.

6.

Program the output voltage to 300 volts.

Output voltage near 300 V.

7.

Record voltage readings at the DVM1 and on the front panel display.

Readings within high voltage

 

 

limits specified in table B-2.

8.

If you are verifying a 3-phase source , repeat steps 1 through 7 for

Readings within specified High

 

phases 2 and 3. Press Phase Select to select the next phase.

range limits (300 V/1 kHz).

1Multiply the DVM reading by the transformer ratio if a ratio transformer is used.

RMS Current Readback Accuracy

This test verifies the current readback. Use the appropriate current shunt with the accuracy specified in table B-1. Use wire of sufficient size to carry the maximum rated current of the ac source (see table 2-1). If you are verifying a 3-phase source, start by verifying phase 1.

Action

1.Turn off the ac source. Connect the load resistor, current shunt, and the DVM across the current shunt as shown in Figure B-1. Use the

following load resistor values:

Agilent 6814B = 7.5Ω; Agilent 6834B = 15Ω; Agilent 6843A = 5Ω

2.Turn on the ac source. In the Output menu, execute the *RST command to reset the unit to its factory default state.

3.Program the output voltage to 100 volts and set the current limit as follows:

Agilent 6814B = 10 A; Agilent 6834B = 5A; Agilent 6843A = 15A Then enable the output by pressing Output On/Off.

4.Record the DVM voltage reading and calculate the rms current. Divide the DVM reading by the current monitor resistor value. Record the front panel reading.

5.If you are verifying a 3-phase source , repeat steps 1 through 4 for phases 2 and 3. Press Phase Select to select the next phase.

Normal Result

*RST

CCannunciator on. Output current near 10 A for Agilent 6814B near 5 A for Agilent 6834B near 15A for Agilent 6843A

Difference between the measured output current and front panel readings are within specified limits.

61

Image 61
Contents User’s Guide AC Power Solutions Agilent Models 6814B, 6834B, and 6843ACertification Warranty InformationGeneral Safety SummarySafety Symbols EMC DeclarationPrinting History Acoustic Noise InformationTable of Contents Performing the Calibration Procedure Specifications Supplemental CharacteristicsError Number List Entry Keys Examples of Front Panel ProgrammingTopic Document OrientationLocation Options, Accessories, and User Replaceable Parts Safety ConsiderationsOption Description Agilent Part NumberCapabilities DescriptionModel Description Front Panel/Remote Operation Output CharacteristicRanges Output VA Capability AC Source Output Characteristic in real-time modePage Inspection CleaningDamage Packaging Material Items SuppliedBench Operation LocationRack Mounting Not block the fan exhaust at the rear of the unitInput Connections Installing the Power CordInput Source and Line Fuse Output Connections Output ConnectionsVoltage Drops Wire ConsiderationsCurrent Ratings Ampacity and Resistance of Stranded Copper ConductorsRemote Sense Connections Remote Sense ConnectionsDigital Connections Trigger ConnectionsOVP Considerations Output RatingController Connections Gpib ConnectorRS-232 Connector Interface CommandsRS-232 Interface RS-232 Data FormatNull Modem Interface Lines Response Data Terminator Hardware HandshakePreliminary Checkout IntroductionCheckout Procedure Using the KeypadVolt Procedure Display ExplanationProtclear Currlev VoltprotCurrprot Currprot OFFCase of Trouble Error MessagesLine Fuse Page Front Panel, Overall View Front Panel DescriptionAC+DC System Keys Display Command FunctionFunction Keys Annunciator On Phase SelectedImmediate Action Keys Meter Display Keys Scrolling KeysDisplay Measurement KBESSEL, RectDisplay Output Control KeysRST CLS Protection and Status Control KeysInitimmed Trigger and List Control KeysIMM AbortThrough Are used for entering numeric values Is the decimal Entry KeysProcedure for Single-Phase AC Sources Setting the Output Voltage AmplitudeExamples of Front Panel Programming Set the output to 120 V rms as followsPhase 3 to 235 Vrms Procedure for Three-Phase AC SourcesAction Display To verify the output, you can measure it as followsSetting a Protection Feature Setting the Output FrequencyOvercurrent protection feature as follows Step Transient Using Transient Voltage ModesVoltt Voltm StepVoltm Pulse Pulse TransientWidth DcycleAction List TransientVoltm Fixed Voltm List Count Volt EOL Step Auto Trigger Delays and Phase SynchronizationExample Display Syncsour PhaseSyncphas Initimmed Example Voltm Step Syncsour Phase Syncphas InitimmedProgramming Slew Rates Using Slew Rates to Generate WaveformsSlew Voltm FixedStep Slewm SlewtMeasuring Peak Inrush Current Syncsour Phase SyncphasRange CurrpeakTo configure the RS-232 interface, proceed as follows Setting the Gpib Address and RS-232 ParametersSaving and Recalling Operating States Action Display To set the Gpib address, proceed as followsAddress Noutputs Table A-1. Performance Specifications1 SpecificationsAgilent 6814B Agilent 6834B Agilent 6843A Table A-2. Supplemental Characteristics Supplemental CharacteristicsSpecifications a Page Equipment Required CharacteristicsRecommended Model Turn-On Checkout Procedure Test SetupPerforming the Verification Tests RMS Current Readback Accuracy Voltage Programming and Measurement AccuracyAction Normal Result Agilent 6814B Current Measurement Accuracy Performing the Calibration ProcedureAgilent 6834B Current Measurement Accuracy Agilent 6843A Current Measurement AccuracyFront Panel Calibration Enable Calibration ModeFront Panel Calibration Menu Calibrating the OVP trip point Calibrating and Entering Voltage Calibration ValuesCalibrating and Entering Current Calibration Values Calibrating the Output Impedance Agilent 6843A only Changing the Calibration PasswordSaving the Calibration Constants CalpassTable B-3. Gpib Calibration Error Messages Calibration Error MessagesCalibration Over the Gpib Agilent Calibration Program ListingFigure B-2. Calibration Program Listing Sheet 1 Figure B-2. Calibration Program Listing Sheet 2 Page Table C-1. Error Numbers Error Number ListError Messages Error Messages C Page Index Index Index Canada Australia/New Zealand United States Latin AmericaEurope Asia Pacific JapanManual Updates
Related manuals
Manual 20 pages 27.01 Kb Manual 187 pages 38.86 Kb

6834B, 6843A, 6814B specifications

Agilent Technologies, a leader in electronic test and measurement equipment, offers a range of powerful signal sources including the 6843A, 6834B, 6814B, 6813B, and 6811B models. These instruments are designed to support various applications in research, development, and manufacturing, providing precise signal generation capabilities.

The Agilent 6843A is a versatile signal generator known for its exceptional frequency range and modulation capabilities. It supports an extensive bandwidth, making it ideal for applications that require high-frequency signal generation. With its superior phase noise performance, the 6843A is an excellent choice for radar, wireless communications, and electronic warfare applications. The instrument features an intuitive user interface, allowing engineers to set parameters quickly and efficiently.

Next, the Agilent 6834B offers exceptional performance characteristics, including high output power and low distortion. This signal generator is particularly noted for its ability to produce complex modulation formats, making it suitable for testing advanced wireless communication systems. With a reliable and stable output, the 6834B ensures accurate and repeatable measurements, which is vital for thorough testing processes.

The 6814B model stands out for its dual-channel capabilities, allowing users to generate simultaneous signals for testing multiple components or systems. This feature significantly enhances testing efficiency and flexibility for engineers. With built-in arbitrary waveform functionality, users can create custom waveforms, making the 6814B suitable for a wide range of applications including device characterization and signal processing research.

For those seeking a more compact solution, the Agilent 6813B provides essential signal generation features without compromising on performance. It is designed for a variety of applications across telecommunications and consumer electronics, featuring a straightforward interface and robust performance metrics.

Lastly, the 6811B is an entry-level yet capable model that supports a broad spectrum of testing needs. Perfect for educational and laboratory environments, it provides essential functionalities required for effective signal generation and analysis.

Overall, Agilent Technologies' 6843A, 6834B, 6814B, 6813B, and 6811B signal generators offer an array of features and technologies that cater to various application needs. Their precision, reliability, and user-oriented designs position them as invaluable assets in any testing environment, ensuring engineers can carry out their work with confidence and accuracy.