Quick Installation Guide
Rosemount 2051
Operation and Maintenance
Proof Test and Inspection
The following proof tests are recommended. Proof test results and corrective actions taken must be documented at www.emersonprocess.com/rosemount/safety/certtechdocumentation.htm in the event that an error is found in the safety functionality.
Use "Table 1: Field Communicator Fast Key Sequence" to perform a Loop Test, Analog Output Trim, or Sensor Trim. See the 2051 reference manual
Proof Test
This proof test will detect 99% of DU failures not detected by the 2051 automatic diagnostics.
1.Conduct a Loop Test. On HART host/communicator enter the Fast Key Sequence 1, 2, 2.
a.Enter the milliampere value representing a high alarm state
b.Check the reference meter to verify the mA output corresponds to the entered value.
c.Enter the milliampere value representing a low alarm state
d.Check the reference meter to verify the mA output corresponds to the entered value.
2.Perform a minimum two point sensor calibration check using the
a.If necessary, use one of the “Trim” procedures available in the 2051 reference manual to calibrate.
NOTE
The user determines the
Visual Inspection
Not required.
Special Tools
Not required
Product Repair
All failures detected by the transmitter diagnostics or by the
The 2051 is repairable by major component replacement. Follow the instructions in the 2051 reference manual (document number
Reference
Specifications
The 2051 must be operated in accordance to the functional and performance specifications provided in the 2051 reference manual.
Failure Rate Data
The FMEDA report includes failure rates and common cause Beta factor estimates. This report is available at www.emersonprocess.com/rosemount.
15