STEP 1 CONTINUED...

Figure 2. Torque values for the flange and flange adapter bolts

 

Bolt Material

Head Markings

Initial Torque Final Torque

Carbon Steel (CS)

Stainless Steel (SST)

316

316 R

 

 

300 in.-lbs.

650 in.-lbs.

B7M

 

 

 

150 in.-lbs.

300 in.-lbs.

 

 

B8M

316

STM SW

316 316

D. O-rings with Flange Adapters

WARNING

Failure to install proper flange adapter O-rings may cause process leaks, which can result in death or serious injury. The two flange adapters are distinguished by unique O-ring grooves. Only use the O-ring that is designed for its specific flange adapter, as shown below.

Rosemount 3051S / 3051 / 2051 / 3095

Flange Adapter

O-ring

PTFE Based

Elastomer

Rosemount 1151

Flange Adapter

O-ring

PTFE

Elastomer

Whenever the flanges or adapters are removed, visually inspect the o-rings. Replace them if there are any signs of damage, such as nicks or cuts. If you replace the o-rings, re-torque the flange bolts and alignment screws after installation to compensate for seating of the PTFE o-ring.

6

Image 6
Contents Quick Installation Guide Rosemount 2051CF Series Flowmeter TransmitterElectrical shock can result in death or serious injury Explosions could result in death or serious injuryProcess leaks may cause harm or result in death Conduit/Cable EntriesMount the Transmitter ApplicationsFlow Optional Mounting Brackets Bolting Considerations Common Transmitter AssembliesRing Rings with Flange AdaptersFlange Adapter Ptfe BasedConsider Housing Rotation Inline Gage Transmitter OrientationSET the Jumpers MA Hart Connect the Wiring and Power RL ≥ 250ΩWiring Power Supply for 4-20 mA Hart Power Supply for 1-5 Vdc Hart Low PowerVerify Configuration Function MA HartTrim the Transmitter Using the Field CommunicatorUsing the Transmitter Zero Adjustment Buttons Zero TrimSafety Instrumented Systems InstallationConfiguration Operation and Maintenance ReferenceApproved Manufacturing Locations Safety Failure ValuesProduct Certifications European Directive InformationNorth American Certifications Hart ProtocolHazardous Locations Certifications European CertificationsRosemount IECEx Certifications Tiis Certifications Inmetro CertificationsCCoE Certifications E4 and I4 combination Combinations of CertificationsE1, I1, N1, and ND combination E5 and I5 combinationRosemount Rosemount Rosemount