Eagle Home Products Eagle Series instruction manual Stel alarm toxics only

Page 21

STEL alarm (toxics only)

If a toxic gas channel’s average gas reading for the past 15 minutes exceeds the STEL alarm setting:

STEL displays in the alarm field for that channel.

The channel’s display line flashes.

The buzzer sounds a pulsed tone.

The alarm lights flash.

TWA alarm (toxics only)

If a toxic gas channel’s average gas reading for the past 8 hours exceeds the TWA alarm setting:

TWA displays in the alarm field for that channel.

The channel’s display line flashes.

The buzzer sounds a pulsed tone.

The alarm lights flash.

Over range alarm

If a channel’s gas reading exceeds that channel’s full-scale setting:

OVER displays in the alarm field for that channel.

The channel’s display line flashes.

The buzzer sounds a pulsed tone.

The alarm lights flash.

Eagle Instruction Manual

Alarms • 15

Image 21 Contents
RKI Instruments, Inc Whipple Rd. Hayward, CA 94544 510 Portable Multi-Gas DetectorWarranty WarrantyTable of Contents Table of Contents Display ModeCalibration Appendix F Eagle Tank Tester Model Introduction OverviewIntroduction About this Manual Specifications SpecificationsEagle Specifications Standard Sensor Specifications Combustible OxygenCarbon Case DescriptionControl Panel ButtonsBattery Charger Connector Alarm LightsInterface Port BuzzerSensors Methane Elimination Switch CAL/SETUP SwitchCircuit Boards Operation Starting Up the EagleOperation Operation Monitoring Combustible Gas in the PPM Range Normal OperationFull Response Mode Conversion Factors Methane Calibration Monitoring Combustible Gases Other than MethaneTarget Gas LEL Factor Turning Off the Eagle Setting User AccessAlarms Alarm IndicationsAlarms Stel alarm toxics only Low flow alarm Sensor failure alarm and emergency operation Resetting Gas Alarms Display Mode Display ModeUser and Station ID Screen Peak Screen Elapsed Time ScreenTo enter a user and station ID Battery Voltage Screen TWA/STEL ScreenDate/Time Screen Clear Data Logger Screens Remaining Log Time ScreenTo clear the data log Setup Mode Tips for Using Setup ModeSetup Mode Updating the Battery Type Setting Entering Setup ModeUpdating Channel Settings Setting a custom target gas label Target Gas Fullscale PPM Readings Equivalent to 100% LELUpdating oxygen channel settings Updating toxics channel settings Setting a custom target gas label Updating Combustible Gas Channel Units of Measure Updating the Alarm Point Settings Updating the Eagle’s Serial Number Updating the Lunch Break SettingUpdating the Alarm Silence Setting Updating the Alarm Latching SettingTurning the User ID Function On or Off Updating the Auto Calibration SettingsTurning the Auto Fresh Air Function On or Off Updating the Back Light SettingUpdating Log Data Over Write Setting data log option Updating the Interval Time Setting data log optionUpdating the Date and Time Settings data log option Updating the Time Calibration Setting data log optionUpdating the Zero Follow Settings Updating the Confirmation Beep SettingReturning to Default Settings To reset all default settings To reset all default alarm point settingsTo reset the oxygen zero setting Calibration Supplies and Equipment CalibrationPreparing for Calibration CalibrationCalibrating the Eagle Calibration Calibrating with the Single Calibration method Calibration Displaying the Battery Voltage MaintenanceTo display the battery voltage Replacing Alkaline BatteriesTo replace the combustibles sensor Replacing SensorsTo replace the oxygen sensor To replace the H2S or CO sensor Parts List Appendix a Parts ListAppendix a Parts List Part Number DescriptionPart Number Appendix B Methane Elimination Appendix B Methane EliminationMonitoring combustible gases other than hexane Specifications Appendix C Non-Standard Toxic Gas SensorsDescription Non-Standard Toxic Gas Sensors SpecificationsCalibrating Non-Standard Toxic Gas Sensors Adjusting the sensor controls Replacing Non-Standard Toxic Gas Sensors Parts List Non-Standard Toxic Gas Sensors Parts ListAppendix D Carbon Dioxide Sensors Alarm TWA Alarm Stel AlarmCarbon Dioxide Sensor Specifications Appendix D Carbon Dioxide SensorsDemand Zero for Carbon Dioxide Sensors Normal Operation of Carbon Dioxide SensorsCalibrating Carbon Dioxide Sensors Carbon Dioxide Fresh Air ReadingsAdjusting the sensor controls Replacing Carbon Dioxide Sensors Parts List Carbon Dioxide Sensors Appendix E Infrared Methane Sensors Range Alarm TWA Alarm Stel AlarmTarget Gases Infrared Methane Sensor SpecificationsSensor Analog PCB Socket Calibrating Infrared Methane SensorsReplacing Infrared Methane Sensors Parts List Infrared Methane SensorsAppendix F Eagle Tank Tester Model To use the float probe assemblyTo eliminate an oxygen alarm that is not being used AlarmsTo attach the dilution fitting Appendix F Eagle Tank Tester ModelCalibration Parts List Eagle Tank Tester ModelAppendix G Five-Gas and Six-Gas Models Displaying Additional ChannelsAppendix G Five-Gas and Six-Gas Models Calibration, Display, and Setup Modes Appendix H Eagle Transformer Gas Tester Model Appendix H Eagle Transformer Gas Tester ModelAlarms Appendix I Installing the Data Logger Board Appendix I Installing the Data Logger Board