Eagle Home Products Eagle Series instruction manual Calibration

Page 49

3.To adjust the values on the screen, hold down the SHIFT/button, and press the DISP/ADJ button. The Auto Calibration screen for the combustible gas channel displays.

4.Use the AIR/(increase) and SHIFT/(decrease) buttons to set the correct combustible gas value.

5.Press the POWER/ENTER button to enter the new setting. The Auto Calibration screen for the next channel displays.

6.Repeat steps 4 and 5 to set the correct values for the remaining channels and return to the Calibration Values screen.

NOTE: The RKI Four Gas Cylinder contains approximately 12% O2

by volume. Be sure to set the “OXY” reading to agree with the concentration listed on the cylinder’s label, not zero.

7.With the Calibration Values screen displayed, press the POWER/ENTER button. The gas readings flash.

8.Connect the tubing from the regulator to the Eagle’s probe.Wait approximately 1 minute or until the readings stabilize.

9.Press the POWER/ENTER button to set the calibration to the programmed values.

If a sensor(s) cannot calibrate to the proper value, FAIL PUSH AIR KEY displays and the Eagle lists the sensor(s) that failed to calibrate. (The other sensors calibrate normally.) The buzzer and alarm lights activate. Press the AIR/button to reset the alarm and return to the Calibration menu. Replace the failed sensor(s), then repeat calibration.

10.AUTO CALIBRATION END displays, then the Calibration menu displays.

11.Disconnect the tubing from the probe.

12.Unscrew the regulator from the calibration cylinder.

13.Press the SHIFT/button to place the prompt next to the NORMAL OPERATION menu option, then press the POWER/ENTER button to return to the normal screen.

Eagle Instruction Manual

Calibration • 43

Image 49 Contents
RKI Instruments, Inc Whipple Rd. Hayward, CA 94544 510 Portable Multi-Gas DetectorWarranty WarrantyTable of Contents Table of Contents Display ModeCalibration Appendix F Eagle Tank Tester Model Overview IntroductionIntroduction About this Manual Specifications SpecificationsEagle Specifications Combustible Oxygen Standard Sensor SpecificationsCarbon Case DescriptionControl Panel ButtonsBattery Charger Connector Alarm LightsInterface Port BuzzerSensors CAL/SETUP Switch Methane Elimination SwitchCircuit Boards Starting Up the Eagle OperationOperation Operation Monitoring Combustible Gas in the PPM Range Normal OperationMonitoring Combustible Gases Other than Methane Full Response Mode Conversion Factors Methane CalibrationTarget Gas LEL Factor Turning Off the Eagle Setting User AccessAlarm Indications AlarmsAlarms Stel alarm toxics only Low flow alarm Sensor failure alarm and emergency operation Resetting Gas Alarms Display Mode Display ModeUser and Station ID Screen Elapsed Time Screen Peak ScreenTo enter a user and station ID TWA/STEL Screen Battery Voltage ScreenDate/Time Screen Remaining Log Time Screen Clear Data Logger ScreensTo clear the data log Tips for Using Setup Mode Setup ModeSetup Mode Updating the Battery Type Setting Entering Setup ModeUpdating Channel Settings Setting a custom target gas label Target Gas Fullscale PPM Readings Equivalent to 100% LELUpdating oxygen channel settings Updating toxics channel settings Setting a custom target gas label Updating Combustible Gas Channel Units of Measure Updating the Alarm Point Settings Updating the Eagle’s Serial Number Updating the Lunch Break SettingUpdating the Alarm Silence Setting Updating the Alarm Latching SettingTurning the User ID Function On or Off Updating the Auto Calibration SettingsTurning the Auto Fresh Air Function On or Off Updating the Back Light SettingUpdating Log Data Over Write Setting data log option Updating the Interval Time Setting data log optionUpdating the Date and Time Settings data log option Updating the Time Calibration Setting data log option Updating the Confirmation Beep Setting Updating the Zero Follow Settings Returning to Default Settings To reset all default alarm point settings To reset all default settingsTo reset the oxygen zero setting Calibration Supplies and Equipment CalibrationPreparing for Calibration CalibrationCalibrating the Eagle Calibration Calibrating with the Single Calibration method Calibration Displaying the Battery Voltage MaintenanceTo display the battery voltage Replacing Alkaline BatteriesTo replace the combustibles sensor Replacing SensorsTo replace the oxygen sensor To replace the H2S or CO sensor Parts List Appendix a Parts ListAppendix a Parts List Part Number DescriptionPart Number Appendix B Methane Elimination Appendix B Methane EliminationMonitoring combustible gases other than hexane Specifications Appendix C Non-Standard Toxic Gas SensorsDescription Non-Standard Toxic Gas Sensors SpecificationsCalibrating Non-Standard Toxic Gas Sensors Adjusting the sensor controls Replacing Non-Standard Toxic Gas Sensors Parts List Non-Standard Toxic Gas Sensors Parts ListAppendix D Carbon Dioxide Sensors Alarm TWA Alarm Stel AlarmCarbon Dioxide Sensor Specifications Appendix D Carbon Dioxide SensorsDemand Zero for Carbon Dioxide Sensors Normal Operation of Carbon Dioxide SensorsCalibrating Carbon Dioxide Sensors Carbon Dioxide Fresh Air ReadingsAdjusting the sensor controls Replacing Carbon Dioxide Sensors Parts List Carbon Dioxide Sensors Appendix E Infrared Methane Sensors Range Alarm TWA Alarm Stel AlarmTarget Gases Infrared Methane Sensor SpecificationsSensor Analog PCB Socket Calibrating Infrared Methane SensorsReplacing Infrared Methane Sensors Parts List Infrared Methane SensorsAppendix F Eagle Tank Tester Model To use the float probe assemblyTo eliminate an oxygen alarm that is not being used AlarmsTo attach the dilution fitting Appendix F Eagle Tank Tester ModelCalibration Parts List Eagle Tank Tester ModelDisplaying Additional Channels Appendix G Five-Gas and Six-Gas ModelsAppendix G Five-Gas and Six-Gas Models Calibration, Display, and Setup Modes Appendix H Eagle Transformer Gas Tester Model Appendix H Eagle Transformer Gas Tester ModelAlarms Appendix I Installing the Data Logger Board Appendix I Installing the Data Logger Board